Oxygen Monitoring
11
11-9
Part Number: 4116841-001 Rev: E
Narkomed MRI-2 Operator’s Manual
Oxygen Alarm Messages
The following list contains all HI, ME, and LO alarms associated with
oxygen monitoring.
%OXYGEN
LOW (HI)
The Narkomed MRI-2 continuously compares the current inspiratory
oxygen percentage with the preset low oxygen alarm limit. If the measured
oxygen concentration falls below the low alarm limit, the HI message
%OXYGEN LOW
appears in the Alarm window, and a continuous audible
alarm sounds.
%OXYGEN
HIGH (LO)
If the measured inspiratory oxygen concentration exceeds the preset high
alarm limit, the LO message
%OXYGEN HIGH
appears in the Alarm
window, and a single-tone audible alarm sounds.
O2 SENS
DISC (LO)
If the oxygen sensor cord becomes disconnected (or is damaged enough to
cause an open circuit), the LO message
O2 SENS DISC
appears in the
Alarm window, and a single-tone audible alarm sounds.
O2 CAL DUE
(LO)
The LO message
O2 CAL DUE
appears in the Alarm window when the
oxygen sensor needs to be calibrated (more than 18 hours have elapsed
since the last calibration).
During oxygen sensor calibration, the Narkomed MRI-2 checks the
sensor's output against a range of acceptable output voltages. There are
three possible causes for deviation from within this range.
•
Exhausted sensor.
If the sensor's capacity is exhausted, its output
voltage will not meet the required minimum.
•
Incorrect calibration environment.
If the sensor is exposed to an
excessive oxygen during calibration, the sensor's output will be above
or below the acceptable output range.
•
Improper waiting.
If the proper waiting period is not allowed for a new
sensor or for a sensor removed from the sensor housing, the sensor's
output may be above or below the acceptable output range.
If a sensor error condition is detected during monitoring, the LO message
O2 CAL DUE, O2 CAL ERR, or O2 NOT CAL
appears in the Alarm window
and operation continues. Try to recalibrate the sensor; if the message
remains, replace the sensor cell.
O2 CAL ERR
The LO message
O2 CAL ERR
appears in the Alarm window when the
oxygen percentage is greater than 103%.
If a sensor error condition is detected during monitoring, the LO message
O2 CAL DUE, O2 CAL ERR, or O2 NOT CAL
appears in the Alarm window
and operation continues. Try to recalibrate the sensor; if the message
remains, replace the sensor cell.
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...