14
Routine Maintenance and Cleaning
14-22
Part Number: 4116841-001
Narkomed MRI-2 Operator’s Manual
Rev: E
Cleaning and
Disinfecting
the Vaporizer
Exterior
Clean the vaporizer with a soft lint-free cloth moistened with mild
detergent and water. Follow up with a 70% to 90% diluted solution of ethyl
or isopropyl alcohol or sodium hypochlorite (5.2% household bleach) at
1:500 dilution (100 ppm chlorine). Do not sterilize the vaporizer.
WARNING:
Water and other liquids (with the exception of the
appropriate anesthetic) that enter the vaporizer chamber can
cause injury to the patient or form corrosive products that
affect the function of the vaporizer.
CAUTION:
Do not immerse the vaporizer or allow water or any other
liquid to enter the fill or drain ports or fresh gas inlet or
outlet ports. Any vaporizer suspected of contamination with
water or any other liquid must be returned to DrägerService
for a complete overhaul.
Additional care must be taken with halothane vaporizers. Halothane
contains a stabilization additive called thymol, which evaporates more
slowly than halothane, and collects in the vaporizer. Over time, thymol
may decompose into compounds that affect the wick material and turn the
halothane yellow.
If you see particles in the sight glass of a halothane vaporizer, or if the
halothane turns yellow, rinse the vaporizer chamber with fresh halothane
as follows:
1. Drain the discolored halothane from the vaporizer.
2. Fill the vaporizer with fresh halothane up to the maximum level, then
drain completely.
3. Dispose of the drained halothane in accordance with standard
practices at your facility.
For information about filling and draining the vaporizer, see the
appropriate separate manual supplied with vaporizers that are available
for use with the Narkomed MRI-2.
Cleaning and
Disinfecting
the Open
Reservoir
Scavenger
The scavenger interface does not normally need to be disinfected. However,
if the user facility requires disinfection, an EtO process can be used after
cleaning. Refer back to the cleaning instructions provided under the
“Routine Maintenance“ section in this chapter.
CAUTION:
Do not autoclave the open reservoir scavenger. The
scavenger's flowmeter cannot withstand the heat of
autoclaving.
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...