3
Daily Checkout
3-10
Part Number: 4116841-001
Narkomed MRI-2 Operator’s Manual
Rev: E
Manual and
Automatic
Ventilation
Systems
26. * Check for proper pressure and flow at the Y-piece during the
inspiratory and expiratory phases. Turn the SYSTEM POWER switch
and ventilator power switch to their ON positions. Place the manual/
automatic selector valve in the AUTO position. Adjust the oxygen flow
control valve to a 3 l/min flow. Set the ventilator frequency to 3 BPM,
the I:E ratio to 1:2, and the tidal volume to about 1 liter.
Adjust the ventilator flow control to the maximum of the “low” zone on
the flow gauge. Occlude the patient side of the Y-piece. Fill the
ventilator bellows by pressing the oxygen flush button. Observe the
breathing system pressure gauge as the ventilator cycles.
The pressure gauge must indicate a pressure over 30 cmH
2
O when the
bellows completes its downward travel. The pressure should not exceed
3 cmH
2
O at the end of the expiratory phase, when the bellows
completes its upward travel.
Monitor
27. Check the alarm limit settings. The monitor alarm limits are
automatically set to a default configuration when the
SYSTEM POWER
switch is turned on. Check these settings and adjust them if necessary.
Alarm limits can be adjusted at the beginning of or during a procedure.
Also, make sure that any external monitors (if any) are connected
properly.
28. * Test the alarm functions. Simulate alarm conditions and check for
appropriate alarm signals.
System Flush
29. Flush the system for at least one minute with 100% oxygen by pressing
the oxygen flush button.
Final Position
30. When the end of the daily checkout procedure is complete, verify that:
a. vaporizers are off (handwheels set to 0)
b. the APL valve is open (fully counterclockwise)
c.
the manual/automatic switch is set to BAG
d. all flowmeters indicate 0 (or minimum)
e. the patient suction level is adequate
f.
the breathing system is ready to use with the bag in place and all
hoses connected properly.
WARNING:
Always lock the casters after this anesthesia machine has
been positioned in the MRI scanner room. Magnetic
attractive forces between the magnet and the anesthesia
machine may cause unintentional movement of the
anesthesia machine if the casters are unlocked.
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...