2
General Description
2-28
Part Number: 4116841-001
Narkomed MRI-2 Operator’s Manual
Rev: E
NOTE:
If a HI or ME message is removed from the Alarm window before
you have a chance to read it, you can view the message in the
Alarm Log. The Alarm Log is a separate screen that displays a
sequential record of up to 100 of the most recent HI and ME
messages. For more information, see “Using the Alarm Log” on
page 10-8.
Alarm
Annunciation
Each alarm category is associated with a specific audible signal.
•
HI
- a five-pulse tone pattern that is initially repeated every few
seconds until the alarm condition is resolved.
•
ME
- a three-pulse tone pattern that is repeated every 30 seconds.
•
LO
- a single tone or no sound at all, depending on the urgency of the
alert.
Only the highest-priority, currently active alarm condition is annunciated.
Tones for lower-priority alarm conditions are temporarily suppressed to
minimize the confusion caused by simultaneous alarms.
Ventilation
Alarms
When the system power switch is turned from
STANDBY
to
ON
, the
volume and pressure apnea alarms default to Standby to allow machine
setup without nuisance alarms. An interlock with the ventilator ensures
that when the ventilator is turned on, the alarms are enabled. You can also
enable the alarms individually using the keypad.
When the ventilator is turned off, the following events occur:
•
If the pressure apnea threshold was greater than 15 cm H
2
O when the
ventilator was turned off, the threshold setting is changed to 15 cm
H
2
O. (If the pressure apnea threshold was less than 15 cm H
2
O when
the ventilator was turned off, the threshold retains its setting.)
•
The ME and HI alarms associated with apnea alarms change from
activation after 15 and 30 seconds of apnea to 30 and 60 seconds,
respectively.
When the ventilator is turned back on, the pressure apnea threshold is
restored to its previous set value and the apnea alarms revert to activation
after 15 seconds (ME) and 30 seconds (HI) of apnea.
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...