Routine Maintenance and Cleaning
14
14-11
Part Number: 4116841-001 Rev: E
Narkomed MRI-2 Operator’s Manual
Removing Parts for Cleaning and Disinfection
1. Turn the SYSTEM POWER switch to STANDBY.
2. Disconnect the 22 mm breathing circuit hoses between the ventilator
and the absorber and from the inspiratory valve and the expiratory
hose terminal on the ultrasonic flow sensor. Remove the Y-piece, mask,
and mask elbow from the hoses.
3. Disconnect the 19 mm scavenger hoses connecting the APL valve to the
absorber pole and the ventilator relief valve to the scavenger.
NOTE:
The scavenger interface generally does not need disinfection.
However, if your facility requires a disinfection process for the
scavenger interface and hoses, remove the scavenger and all 19
mm hoses. Refer back to the cleaning instructions under “Open
Reservoir Scavenger Maintenance” or “Scavenger Interface for
Passive Systems Maintenance” for disassembly instructions.
4. Remove the oxygen sensor assembly from the inspiratory valve, and
disconnect the cable that connects the sensor to the monitor.
5. Remove the connector hose assembly between the flow sensor and the
expiratory valve by turning the ring nut counterclockwise.
6. Lift the flow sensor off the bracket.
7. Disconnect the fresh gas hose from the fresh gas outlet.
8. Disconnect the breathing pressure pilot line from the absorber and the
monitor.
9. Remove the absorbent canisters and dust cup. Discard the absorbent.
Refer back to “Replacing the Absorbent” for instructions.
WARNING:
Absorbent is caustic and a strong irritant to the eyes, skin,
and respiratory tract. When removing the absorbent, take
care not to spill its caustic contents.
10. Remove the inspiratory and expiratory valves by turning their ring
nuts counterclockwise until they are free of the system.
11. Remove the ultrasonic flow sensor by turning its ring nut
counterclockwise until it is free of the absorber dome.
12. Remove the pressure gauge by turning the knurled ring nut that
connects the pressure gauge to the gauge mount counterclockwise. Lift
the pressure gauge assembly off the gauge mount. Do not misplace the
o-ring.
CAUTION:
Only the pressure gauge supplied with a Narkomed MRI-
2 should be used.
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...