Daily Checkout
3
3-3
Part Number: 4116841-001 Rev: E
Narkomed MRI-2 Operator’s Manual
7. * Check the battery power. Press the BATTERY TEST button on the
main switch panel. The green indicator must light. The yellow
BATTERY LOW indicator must remain unlit.
NOTE:
This test assumes the anesthesia machine has been plugged in
for 16 hours. The battery charging system works only when the
machine is plugged into an active AC power source. The
charging system takes about 16 hours to charge a fully
discharged battery.
Emergency
Ventilation
Equipment
8. * Verify that backup ventilation equipment is available and functional.
High
Pressure
System
Verification
9. * Check the oxygen cylinder supplies.
a. Disconnect all pipeline gas supply hoses and drain the system.
b. Close the oxygen cylinder valve and remove the cylinder from the
yoke. Verify that there is only one cylinder gasket and there are two
index pins. Verify that the cylinder matches the yoke label. Replace
the cylinder.
c.
Open the oxygen cylinder and check the cylinder pressure gauge. A
full oxygen cylinder registers a pressure of about 1900 psi. Replace
any cylinder with pressure less than 1000 psi. To check for a high
pressure leak, close the cylinder and observe the cylinder pressure
gauge for a prominent decrease in the pressure. With the oxygen
cylinder closed, press the oxygen flush button on the front of the
anesthesia machine. Hold the button in until the pressure gauges
indicate no pressure.
10. * Check the nitrous oxide cylinder supplies.
a. Close the nitrous oxide cylinder valve and remove the cylinder from
the yoke. Verify that there is only one cylinder gasket and two
index pins. Verify that the cylinder matches the yoke label. Replace
the cylinder on the yoke.
b. Open the nitrous oxide flow control valve until the nitrous oxide
pipeline and cylinder pressure gauges indicate zero pressure. Open
a nitrous oxide cylinder and check the cylinder pressure gauge. A
full nitrous oxide cylinder registers a pressure of about 745 psi.
Replace any cylinder with a pressure less than 600 psi. To check for
a high pressure leak, close the cylinder and observe the cylinder
pressure gauge for a prominent decrease in the pressure.
The following table shows the full and minimum pressures for E-
size gas cylinders at 70
°
F, 21
°
C.
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...