14
Routine Maintenance and Cleaning
14-20
Part Number: 4116841-001
Narkomed MRI-2 Operator’s Manual
Rev: E
After cleaning the capsule, perform an EtO process at a temperature not
exceeding 50
°
C. Aerate the sensor according to the manufacturer's
instructions.
Cleaning and
Disinfecting
the Oxygen
Sensor
Housing
The oxygen sensor housing can be immersed for cleaning and disinfection.
Use mild detergent and water for cleaning. For disinfection, follow up with
either sodium hypochlorite (5.2% household bleach) l:500 dilution (100
ppm free chlorine) or 70% to 90% ethyl or isopropyl alcohol.
Cleaning and
Disinfecting
the Oxygen
Sensor
Housing
Cover
The oxygen sensor housing cover can be immersed or wiped during
cleaning and disinfection. Use mild detergent and water for cleaning. For
disinfection, follow up with a 2% glutaraldehyde-based formulation, wet
pasteurization at 70
°
C for 30 minutes, or an EtO process.
Cleaning and
Disinfecting
the
Inspiratory
and
Expiratory
Valves
The inspiratory and expiratory valves come into contact with the patient's
breath. Clean these parts with soft lint-free cloth moistened with mild
alkali detergent and water. Dry the parts before reassembling.The valve
assemblies can be immersed in a 2% glutaraldehyde-based solution, wet
pasteurized, gas sterilized, or autoclaved at a temperature not exceeding
121
°
C. The valve disks can also be autoclaved in certain instances, but
should not be autoclaved routinely.
Be careful while handling the valve disks. Improper handling can damage
them. After cleaning or disinfecting the inspiratory or expiratory valve
assembly, verify that the valve disk and all internal pins are undamaged
and that the valve disks are correctly reinstalled.
Cleaning and
Disinfecting
the
Ultrasonic
Flow Sensor
Housing and
Transducers
Wash the flow housing and transducers with mild detergent and water.
Follow with a distilled water rinse. These components can be immersed.
The flow housing and transducers can be autoclaved at a temperature not
exceeding 121
°
C. Follow the autoclave instructions for the process. Allow
the transducers to normalize for 30 minutes under room ambient
conditions before using them.
Cleaning and
Disinfecting
the
Ultrasonic
Flow Sensor
Electronics
Housing and
Cable
Wipe the electronics housing and sensor cable with a clean, soft cloth
moistened with mild detergent and water. Take care not to allow any fluid
to access the interior of the electronics housing.
The electronics housing and sensor cable can be wiped with a clean, soft,
lint-free cloth moistened with a 70% to 90% diluted solution of ethyl or
isopropyl alcohol or sodium hypochlorite (5.2% household bleach) at 1:500
dilution (100 ppm chlorine).
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...