Breathing Pressure Monitoring
13
13-13
Part Number: 4116841-001 Rev: E
Narkomed MRI-2 Operator’s Manual
Turning the Apnea Pressure Alarm Off
You can turn the apnea pressure alarm off only when the ventilator is off.
When the ventilator is on, the alarm is automatically forced on to ensure
notification of problem conditions.
To turn off the apnea pressure alarm, press the Breathing Pressure
Off
key. Audio annunciation of the alarm is disabled and pressure alarm
messages are erased from the display. The LED next to the Breathing
Pressure
Off
key lights to indicate that the apnea pressure alarm is
disabled.
Turning the Apnea Pressure Alarm On
To turn on the apnea pressure alarm, press the Breathing Pressure
On
key.
The LED next to the Breathing Pressure
On
key lights to indicate that the
apnea pressure alarm is enabled.
NOTE:
After power-up, the pressure alarms are in the standby state.
Breathing Pressure Alarm Messages
The following list contains all HI, ME and LO alarms associated with
breathing pressure monitoring.
APNEA-
PRESSURE
(HI/ME)
When the ventilator is on:
If the measured breathing pressure remains below the threshold
pressure alarm limit for more than 15 seconds, the ME message
APNEA-PRES
appears in the Alarm window and an intermittent
audible alarm sounds.
If the breathing pressure remains below the threshold pressure for an
additional 15 seconds (30 seconds total), the ME message
APNEA-
PRES
is upgraded to a HI message in the Alarm window, and a
continuously repeating audible alarm sounds. During the HI condition,
numeric data remains on the display as long as the monitor detects a
peak pressure at least 10 cm H
2
O greater than PEEP pressure. When
this pressure difference drops and remains below 10 cm H
2
O for more
than 60 seconds, the numeric data is cleared.
When the ventilator is off:
The ME condition does not occur until 30 seconds have elapsed; the HI
condition does not occur until 60 seconds have elapsed.
Off
On
Содержание Narkomed MRI-2
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...