LK-400 Operation Manual
Auto Slide Preparation System Extension Module Package
39
*1 The maximum 96 is the System with BS-96 (96-block feeding); (24) if the System with BS-24
(24-block feeding).
*2 The maximum 400 is the System with SS (400-slide feeding) and DU-400 (400-slide dry chamber)
option;other than these the (100) is the maximum.
manual sectioning
mode).
sectioning condition.
1 to 400 (100):
③
Number of the sectioning
condition (*2)
3 to 8:
④
Sectioning thickness.
5 characters:
⑤
Side type.
9009
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PSD receiving error.
Slide storage data.
Time-out or response
error.
No response from the PC for the slide storage
data transmission.
・
Check the LK-400 screen if there is an error.
・
Restart the LK-400 PC application; and log in
again.
・
Shut down the Main Unit power and the
PC;restart the initialization of the Main Unit
and the LK-400 application. Then retry
section preparation with Auto Sectioning
Mode or with Retry Mode.
9010
<Error Message>
Unit: Remote PC
Symptom:
Communication error
%PBC receiving error.
Barcode check error.
Check the sample barcode if there is an error.
The
system will not work using with
noncorresponding
barcode or exceeded numbers of the digits.