LK-400 Operation Manual
Auto Slide Preparation System Extension Module Package
33
7 – 2 – 3 Block setting
The barcode information and input information are not verified under the Manual Input Mode. Make
sure to match the block number and the position number where the block is set on the feeding tray.
⚫
System does not read the sample block barcode with the Manual Input Mode.
⚫
Ensure to match-up with the position number of the sample tray where the
sample is placed and the sample number on the LK-400 screen. Make sure to
correctly coordinate the numbers with the section condition data and/or print
information.
⚫
Ensure to confirm visually that the positioned sample and the slide print are
matched.
!
Warning
Nos.
Match up the
number (Nos.)
Match-up the placing
tier (shelf) and the
position on the tray
with
96-sample
feeding.
Manual Input Mode screen
Sample tray
2. Input each item with the following references.
・
Essential items to input (indicate with the stain
code selection):
Stain name, Glass type, Thickness, pieces
(quantity). (Thickness and pieces are
available to change).
・
Items available to input
Block ID,
Sample No.,Tissue No., Block No.,
Slide No., Department, Priority
*Input the Slide No. in case of entering the
barcode information.