4.3
Verify Measurement Dashboard
Selecting which Verify measurement to view
•
Select “Verify” in the “
Type”
option (top left corner)
•
Select the “
Date
” when the measurement was done
•
Scroll down to the time of the measurement in the “
Date
/
Time
” window.
The Dashboard Content:
Result:
Shows the overall result of the measurement (provided by the
instrument’s internal algorithms). The colour of this field is Green for
successful measurements and Red if there was no match or if any other
failure occurred.
Device ID:
The unique ID code of the instrument that was used for this measurement.
Done by:
Shows the name of the user who performed this measurement
Scan Time:
Shows the exposure time used for the measurement
Status Code:
Shows the status code reported by the instrument
Verified Against:
Shows the name and CAS number of the reference substance that was
selected for this scan.
Mathematical correlation:
Shows the mathematical correlation factor between the measured
spectrum and the spectrum of the selected reference substance. Please
note that the “Pass Limit”, which is set on the instrument, defines the
threshold for what mathematical correlation factor that is to be defined as