21
9. Press RESET to reset the motor positions. The
Actual Motor position should be changed to
zero and the force should be zero or very
close to zero.
MOUNTING ON CHAMBER 1
CH1:
Force 1: -12.1 mN
Actual Motor 1 pos. : +0.0um
ARTERIES
PARAMETER
STOP
MOTOR
RESET
TOGETHER
RELEASE
Wating for command
MOTOR M1 MANUAL:
WARNING: DO NOT PRESS RELEASE AT
THIS STAGE IT WILL DESTROY YOUR
MOUNTED ARTERY.
10. Go into the ARTERIES PARAMETER menu
CH1 ARTERIES PARAMETER
SELECT
SELECT
SELECT
Eyepiece a1: 1.0
Eyepiece a2: 2.0
Norm. Force: 1.0 mN
ENTER
START
NORM.
Segment Length: 0.4 mm
11. Enter the Eyepiece values of the artery mounted
in the given chamber. When the a1 and a2 values
have been entered the 630MA will calculate the
artery Segment Length based on the a1, a2 and
Eyepiece Cal values. Check if the segment length
is approx. correct. The gap in the jaws are 2.00
mm (shown with red arrows in the figure to the
right).
12. Enter the Norm Force. This value is the force that
the vessel is stretched successively at the stretch
steps 1, 2, 3.... The default is 1.0mN but this has
to be lower for small arteries and higher for
larger arteries for a successful normalization. If
the normalization fails it can always be repeated
with new values for the Norm Force.