6. Enter the Norm. Pressure, Norm Factor and
Eyepiece cal values. It is very important that
the Norm Pressure and Norm factor is right
for the type of artery mounted in the chamber.
There can be a huge difference between
artery type and species e.g for rat mesenteric
arteries the Norm Pressure is 13.3kPa and the
Norm Factor is 0.9 but for mouse mesenteric
arteries the Norm Pressure is 13.3kPa and
the Norm Factor 1.1. To find the Norm factor
for your specific artery and specie please
read DMT Normalization Guide.
7. Go into the Mounting Artery
8. Select the appropriate Chamber
MOUNT ON CHAMBER NO.
SELECT
SELECT
SELECT
SELECT CHAMBER 1
ENTER
SELECT
SELECT CHAMBER 2
SELECT CHAMBER 3
SELECT CHAMBER 4
IMPORTANT: The Norm Factor has to be found for the type of vessel and species BEFORE
using the automated normalization.