16
3.2.2 MOUNTING STEP TWO
• Using the manual micropositioner move
the two mounting pins as close together as
possible without touching each other.
• Take the isolated artery/tissue section and
slide it over the two pins as shown in the
figures below.
The artery is now mounted correctly and the
heat of the 630MA interface can be turned ON.
When the buffer in the chamber have reached
37C, replace the buffer in the chamber with fresh
well gassed warm buffer. Wait 5 min and start
the Automated Normalization as described in
Chapter 3.3.5.
NOTE: For large arteries in could be necessary to add a small tension (1-2mN) to the
mounted artery on the pins to avoid that the artery falls of the pins during heating and
replacement of the buffer. Remember to place the pins as close to each other as possible
without touching just before starting the automated normalization.
Содержание 630MA
Страница 1: ...USER GUIDE MYOGRAPH SYSTEM 630MA ...
Страница 2: ...AUTOMATED MULTI WIRE MYOGRAPH SYSTEM ...