IMPORTANT:
In exceptional cases, the supports (jaws or pins) may need to be removed from the
chamber and cleaned individually to assure proper cleaning of all support surfaces.
NEVER SOAK THE SUPPORTS IN ANYTHING STRONGER THAN 8% ACETIC ACID FOR
EXTENDED PERIODS OF TIME (i.e. several hours or overnight)!
4.2 MAINTENANCE OF THE FORCE TRANSDUCER
The force transducer is the most delicate and
fragile component of the Wire Myograph.
Extreme care should be used when handling or
touching the force transducers.
As a part of daily maintenance, inspect the grease
around the transducer pin extending from the
transducer housing pinhole before starting any
experiment (see figure 4.1). Insufficient grease in
this area will allow buffer and water to enter the
transducer housing and causing damage to the
force transducer.
IMPORTANT:
• DMT recommends that the high vacuum grease sealing the transducer pinhole is
checked and sealed at least once a week, especially if the Wire Myograph is used
frequently.
• DMT takes no responsibilities for the use of any other kinds of high vacuum grease
other than the one available from DMT.
• DMT takes no responsibilities for any kind of damage applied to the force transducers.