31
7. Incubate the chamber and supports for
30 minutes with 2mM T-1210 Tetrakis-
(2-pyridylmethyl)-ethylenediamine solution
dissolved in double distilled water.
8. Use a cotton-tip applicator to mechanically
clean all the affected surfaces during the last
15 minutes of the incubation period.
9. Wash the chamber and supports several
times with double distilled water.
10. Incubate the chamber with 96% ethanol for
10 minutes while continuing the mechanical
cleaning with a cotton-tip applicator.
11. Remove the ethanol solution and wash a few
times with double distilled water. Incubate
the chamber and supports with an 8% acetic
acid solution for 10 minutes and continue the
mechanical cleaning with a swab-stick.
12. Wash the chamber and supports several
times with double distilled water.
13. Dry the surfaces using absorbent paper (i.e.
Kim-Wipes) or cotton-tip applicators.
If red or brown discolorations appear on the chamber sides or on the supports, the following cleaning
procedure will work in most cases:
IMPORTANT:
• Be very careful using HCl or HNO
3
because these acids may cause extreme damage to
the stainless steel chambers and supports.
• DO NOT USE bleach to clean the chambers. Repeated use of chlorinated solutions
such as bleach and HCl will cause damage to the stainless steel parts of your myograph
system. Avoid using them if at all possible.
• After cleaning, ALWAYS check that the grease around the transducer pin is sufficient
to keep the buffer and water from entering the transducer housing.
• Any acids spilled outside the high grade steel chamber onto the myograph has to be
removed immediately.