78
| PT5300
example:
:OUTPut:HD5:del 0,1,144.0;
sets delay to 1 line and 144.0 ns
:OUTPut:HD5:del?;
response: +0,+001,+00141.4 (rounded)
:OUTPut:HD5:del -0,-561,-144.0;
sets delay to -561 lines and 144.0 ns
:OUTPut:HD5:del?;
response: -0,-561,-00141.4 (rounded)
:OUTPut:HD5:del -0,-562,0.0;
sets delay to -562 lines and 0.0 ns
:OUTPut:HD5:del?;
response: -0,-562,-00000.
:OUTPut:HD1?
:OUTPut:HD4?
:OUTPut:HD5?
Command to display complete setting of the MSG-OPT-8612 HD Test Pattern Generator. The response contains 8
fields:
<
Pattern
>, <
Text
>, <
System
>, <
Audio signal
>, <
Audio level
>, <
Field
>, <
Line
>, <
HTime
>
OUTPut:LTCG1:FORMat
OUTPut:LTCG2:FORMat
Command to set the format of the LTC generator module. The delay is defined by four parameters:
<
Format
>
, <
Syncmode
>, <
Hour
>, <
Min
>
where
<
Format
>
sets the format to one of the format listed in table 10,
<
Syncmode
>
sets the mode to reset the framecounter (see table 11),
<
Hour
>
sets the sync hour and
<
Min
>
sets the
sync minute. The sync mode is only relevant for 29.97 frames per second, as other formats always stays
synchronized to real time.
24FPS
24 Frames pr second
25FPS
25 Frames pr second
2997NOND 29.97 Frames pr second, non drop
2997DROP 29.97 Frames pr second, drop frame
30FPS
30 Frames pr second
Table 10: LTC Generator formats
NONE No synchronization (Only manual sync.)
CONF Confirm synchronization. Press E to confirm.
AUTO Auto synchronization. No need to do
Table 11: LTC Generator Sync modes
<
Hour
>, <
Min
>
sets the time, when the frame counter shall be resynced. The hour can be set to a number between
0 and 23. The minutes can be set between 0 and 59.
example:
:OUTPut:LTCG1:FORMAT
’24FPS’,’NONE’,0,0;
Sets the format to 24 frames per second, the sync mode and time is not relevant for this mode.
:OUTPut:LTCG1:FORMAT
’24FPS’,’NONE’,0,0;
response: 24FPS,NONE,0,0