PT5300 |
47
Note: Only one phase is needed for audio in PAL environments, due to the simple relation between the audio
sample rate and the PAL system. For audio in NTSC, five different phases are required to be able to synchronize
under all circumstances.
MENU:SDI-TSGx, configure
SUBMNU:<PATTERN> SYSTEM EDH ...
4.8 Reset
4.8.1 Factory Reset
MENU:SDI-TPGX, configure
SUBMNU:<EMB.AUDIO> TIMING ...
The Factory Reset function resets all user-programmed parameters to the factory preset.
This function should not be used except in very unusual situations.
To execute the Factory Reset:
Turn ON the generator while simultaneously pressing the
and
buttons.
After the factory reset is done, a confirming message is shown on the display for 1.5 seconds, and then the instrument
proceeds to the normal startup. The following parameters are reset to factory values:
Parameter:
Factory value:
AUTO ESCAPE
On
PANELLOCK
Off
NORMALLOCK
Off
DIAGNOSE LOCK
Off
RS232-INTERFACE
9600, 8, 1, NONE, XON/XOFF
LOCAL LOCK-OUT
Off (RS232 only)
CONTRAST
17 (middle contrast)
4.8.2 Menu: CONFIG/DIAGNOSE
The diagnose submenu is used to perform internal tests of both the basic instrument and the optional installed
modules. The output signals will be undisturbed during the standard diagnose routines.
SUBMNU:CONFIG/DIAGNOSE, select
E
SELECT:<Main> Options RS232 ...
The tests that will appear in the SELECT menu are:
• Main
• Options
• Display
• Keyboard
• Memory
• Configuration
• Errorqueue