Note!
When setting the target stop to the paper width, make sure that the
respective target or target band is guided without any play but on the other
hand it must not stick.
A target stop which is set too tight can lead to the target becoming distorted,
which, in turn, can falsify the evaluation results.
The transport rollers will start to run and the target will be drawn into the machine as soon
as the target has been recognised by a light barrier on the left side of the evaluation shaft.
Important!
The target must not be pulled or pushed by hand or moved in any other way
once it has been taken up by the machine.
The
DISAG RM IV
is a measuring instrument and as such should be treated
with care!
The following will be shown in the display:
Measureing 1
Target.
The
DISAG RM IV
determines the position of the targets one after the other which means
10 times in the case of an air rifle band and once for single targets.
As soon as this operation has been concluded the following will appear in the display:
Type1234567XXXXX
10 AR-Targets
Now, as a result of this, the new type of target and its number will be placed in the X
position. The cursor will flash above this type of target. Now the seventh type of target has
been set.
If the
ENTER-key is now pressed, the evaluation unit will be set to the new type
of target.