DIESTRO 103APV4R ELECTROLYTE ANALYZER
77 JS Medicina Electrónica
7. TRANSPORTATION OF EQUIPMENT
Perform ISE Cleaning and, if necessary, clean and decontaminate all necessary surfaces of the
equipment.
Wear gloves in all cases.
Prevent damage to the power cord during transport or storage of the
equipment. In case of damage change it for a new one
7.1 Transport within the laboratory field
If the transport does not involve significant movements, it will be enough to verify that the
fastening screws of the Pack container base are well adjusted and carry out the transport
carefully, without tilting or hitting the equipment.
7.2 If the transport involves a more important transfer
But it is reinstalled at most the next day proceed as follows:
1.
Disconnect the pipes of Std. A, Std. B, Lactic Acid (if present) and waste.
2.
Disconnect the analyzer.
3.
Close the bottles of the Pack tightly.
4.
Do not tilt the equipment.
5.
If it is necessary to pack it for transport, use the box in which the equipment was
delivered along with the parts that prevent bumps and movements. If such a box is not
available, use one of similar sizes and fill the sides of the equipment well.
6.
Install by following the installation steps.
7.3 If the equipment is not to be installed in more than two days.
1.
Perform an intensive rinse, if TCO2 option is present also perform a TCO2 rinse.
2.
Disconnect the pipes of Std. A and Std. B. and Lactic Acid (if TCO2 option is installed)
3.
Emptying pipes from the equipment.
a. In the valve menu, open valve A and in manual motion press CW until no more liquids
circulate through the waste pipe.
b. Repeat the operation by opening only valve B.
c. Repeat the operation by opening only the C valve. (if sampler option is installed)
d. Close all valves.
See section:
"MANUAL POSITIONING".
Содержание 103APV4R
Страница 1: ...ELECTROLYTE ANALYZER OPERATOR AND SERVICE MANUAL DIESTRO 103APV4R V 1 0 MARCH 2022 RE 0851 CONTENT ...
Страница 91: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 91 JS Medicina Electrónica Needle sampling with Tip placed ...
Страница 94: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 94 JS Medicina Electrónica 6 Put the needle cover and screw it ...
Страница 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Страница 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Страница 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Страница 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...
Страница 114: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 114 JS Medicina Electrónica Electrode cable Electrode lock Electrode chamber ...