DIESTRO 103APV4R ELECTROLYTE ANALYZER
57 JS Medicina Electrónica
2.3
Using the Left and Right buttons, position the sample so that it is in contact with the
reference electrode and the electrodes for which it reaches volume.
2.4
Once the sample is positioned correctly and without bubbles, lower the lever and press
"Measure".
2.5
The measurement will begin and the Analyzer will continue normally
The operator is responsible for positioning the sample correctly. Position it
from the reference electrode covering as many electrodes as possible with the
amount of sample it counts. Remember that the sample has to be in contact
with the reference electrode. Beware of electrodes that are not in contact with
the sample. The results of such electrodes will be incorrect. It is the
responsibility of the operator to discard the results of such electrodes.
Содержание 103APV4R
Страница 1: ...ELECTROLYTE ANALYZER OPERATOR AND SERVICE MANUAL DIESTRO 103APV4R V 1 0 MARCH 2022 RE 0851 CONTENT ...
Страница 91: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 91 JS Medicina Electrónica Needle sampling with Tip placed ...
Страница 94: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 94 JS Medicina Electrónica 6 Put the needle cover and screw it ...
Страница 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Страница 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Страница 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Страница 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...
Страница 114: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 114 JS Medicina Electrónica Electrode cable Electrode lock Electrode chamber ...