DIESTRO 103APV4R ELECTROLYTE ANALYZER
64
JS Medicina Electrónica
Serial output configuration:
To configure serial output, navigate to the menu:
Route
: "HOME" screen → → Configuration → Connections
Press "Serial Output" to enable it. Select the port from the drop-down menu.
Here you can choose the communication speed and enable/disable the Xon/Xoff protocol.
"Mode" will allow you to configure the Analyzer to validate the data before sending it over the
network:
This parameter includes the following options:
●
Validate:
Each time a measurement is completed, the operator must press the send data
icon to send the data or press the back arrow to exit without sending it.
●
Do Not Validate:
Each time a measurement is completed, it will be sent automatically.
●
Validate on Error:
Each time the measurement is finished, and if there is an error, the
operator must acknowledge the transmission by pressing the send button icon or press
the back arrow to exit without sending it. Error-free measurements are automatically sent
to the network, without validation by the operator.
Note
: When using the Sampler, as long as the LIS output is enabled, all measurements will be
sent without operator validation.
Содержание 103APV4R
Страница 1: ...ELECTROLYTE ANALYZER OPERATOR AND SERVICE MANUAL DIESTRO 103APV4R V 1 0 MARCH 2022 RE 0851 CONTENT ...
Страница 91: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 91 JS Medicina Electrónica Needle sampling with Tip placed ...
Страница 94: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 94 JS Medicina Electrónica 6 Put the needle cover and screw it ...
Страница 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Страница 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Страница 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Страница 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...
Страница 114: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 114 JS Medicina Electrónica Electrode cable Electrode lock Electrode chamber ...