DIESTRO 103APV4R ELECTROLYTE ANALYZER
44
JS Medicina Electrónica
9
– PURGES
To access this menu, navigate to:
Route
: "HOME" screen → → Rinse
1. PURGE STDS
During this process, the pack and equipment pipes are filled with both calibration solutions.
2. PURGE LACTIC
In equipment equipped with a module for measuring TCO2, the lactic acid pipe circuit must be
purged when changing the Lactic Acid bottle.
If the equipment has not been used for a long time you must also perform a purge.
After the purge is complete, a TCO2 wash is carried out automatically.
3. AUTOMATIC PURGE
It is performed automatically when a new pack is installed in the analyzer,
The analyzer will first purge the reagent from the stdB bottle and then the reagent from the stdA
bottle.
When the process is finished, a calibration will automatically begin.
If for some reason the pipes of the Pack are disconnected, it is necessary to
perform a purge manually.
Otherwise, calibration, measurement and/or rinsing will fail
Содержание 103APV4R
Страница 1: ...ELECTROLYTE ANALYZER OPERATOR AND SERVICE MANUAL DIESTRO 103APV4R V 1 0 MARCH 2022 RE 0851 CONTENT ...
Страница 91: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 91 JS Medicina Electrónica Needle sampling with Tip placed ...
Страница 94: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 94 JS Medicina Electrónica 6 Put the needle cover and screw it ...
Страница 109: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 109 JS Medicina Electrónica Pack Rear View Cable Uchip ...
Страница 111: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 111 JS Medicina Electrónica Autosampler Pack ...
Страница 112: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 112 JS Medicina Electrónica Rear view detail ...
Страница 113: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 113 JS Medicina Electrónica Front View with Open Lid ...
Страница 114: ...DIESTRO 103APV4R ELECTROLYTE ANALYZER 114 JS Medicina Electrónica Electrode cable Electrode lock Electrode chamber ...