Title:
DICTUM HEALTH IDM100 MEDICAL TABLET INSTRUCTIONS FOR USE
DCO#:
00271
Document#:
870-00001-01
Originating Author:
E Keate
Modifying Author:
R. Massie
Date Created:
9/04/2018
Date Modified:
05/28/2020
Revision:
AR
Page
152 of 175
To Perform an Overpressure Test in Neonate Mode:
1. Select
Neonatal
.
2.
Connect a manometer, volume, and the hand bulb to the module using “T” adapters.
3. Tap
START
to close the Control Valve and the Dump Valve. Ensure that the Hand Bulb Exhaust Valve
is closed.
4. Increase the pressure to approximately 140 mmHg using the hand bulb. To stop the test at any time,
tap
STOP
.
5. Slowly increase the pressure from 140 mmHg to the overpressure point.