Title:
DICTUM HEALTH IDM100 MEDICAL TABLET INSTRUCTIONS FOR USE
DCO#:
00271
Document#:
870-00001-01
Originating Author:
E Keate
Modifying Author:
R. Massie
Date Created:
9/04/2018
Date Modified:
05/28/2020
Revision:
AR
Page
146 of 175
2. Tap
START
to close the Control Valve and the Dump Valve. Ensure that the Hand Bulb Exhaust Valve
is closed.
For step 3:
•
A hand bulb is preferred to slowly increase the pressure in the pneumatic system.
•
When applying high pressures, carefully increase the pressure at a rate that will not cause
unwanted overpressure errors (300 mmHg).
•
To stop the time at any time, tap
STOP
.
3. Increase the pressure to approximately 250 mmHg using the hand bulb or the pump on the module.
3-Minute Safety Limit:
•
Safety circuitry on the module only allows the pressure in the pneumatic system to remain above
10 mmHg for 180 seconds, unless inhibited by the LK1 jumper.
•
If this safety time limit is exceeded, the valves will open releasing the pressure.
All pneumatic systems have some degree of expansion and rebound, which are affected by several
factors including volume, material types and material durometers.
Test Sequence:
•
The pneumatic system reaches its pressure equilibrium point after 60 seconds.