3. Ensure that each electrode is connected correctly on the patient. After each electrode is successfully
attached, the electrode status turns green on the Electrode Placement screen.
A red electrode status indicates that the corresponding electrode is disconnected.
Before you can continue, properly connect each electrode. When each electrode is connected and
shows a green status, the
START
button is available.
4. Tap
START
to begin the test. The ECG Preview screen appears.
If you are familiar with the ECG Preview screen functions, go to step 6.
5. Use ECG Preview to prepare to record ECG.
When the ECG initiates, the data is plotted in real time.