“FAIL” appears for leak rates outside the acceptable limit.
5. Tap
DONE
to end the test.
21.4 NIBP Overpressure Test
Prepare to perform an Overpressure test:
1. Stage the equipment for the test:
•
Calibrated Manometer (optional
– may use pressure values supplied by NIBP module)
•
Pneumatic “T” Adapters
•
Volume (500 mL rigid bottle)
•
Hand Bulb
•
Connection Tubing
2. Understand the role of volume and cuffs in the test:
•
A 500 mL volume is recommended to prevent exaggerated leak rates with extremely small
volumes.
•
Blood pressure cuffs have large variations in size and materials causing variability in leak test
results. To limit this effect, use only a 500 mL rigid volume.