Page
139 of 175
20
Maintenance
The expected service life of the IDM100 Medical Tablet is 3 years. The expected service life of any accessory
shipped with the IDM100 Medical Tablet is 90 days, unless labeled differently.
Dictum Health will make available upon request, circuit diagrams, component part lists, descriptions,
calibration instructions, or other information to assist service personnel to repair parts of the IDM100 Medical
Tablet that are designated by Dictum Health as repairable by service personnel.
20.1 Inspect the IDM 100 and Peripherals
CAUTIONS
Perform the following inspections daily to ensure patient safety and proper operation:
•
Check for cracked or broken patient cable, patient leads, power cord, communications cables, display,
and enclosure.
•
Check for bent or missing pins on all cables.
•
Check all cable and cord connections; reseat if any connectors are loose.
20.2 Cleaning
CAUTIONS
To prevent the spread of infection, the IDM100 Medical Tablet and patient cables must be kept clean,
especially patient-contacting components.
Do not let soap or water come into contact with th
e device’s connectors or jacks.
Do not attempt to clean the device or the patient cables by submersing them in liquid, autoclaving, or steam
cleaning.
Do not pour alcohol directly on the equipment or soak any components in alcohol. If alcohol or liquids spill
into the device while cleaning, arrange to have it checked before further use. Refer to section 1.1 Warranty
Statement for Dictum Health contact information.
Follow these cleaning instructions monthly, or as needed:
1. Disconnect the power plug from the AC outlet.
2. Wipe the exterior of the device, cables, and accessories with a damp cloth using mild detergent diluted in
water.
3. Use 70 % isopropyl alcohol to disinfect Spiragrip handle, ECG cables, blood pressure cuffs and hoses,
SpO2 sensor and cable, temperature probe and dock, and stethoscope.
4. Dry all components with a clean, soft cloth or paper towel.
5. Wait at least 10 minutes to allow all traces of alcohol to evaporate before powering on the device.
20.3 Changing the Battery