Step 14 - Calibrate Label and Test Cycle
Goal:
Calibrate the label gap sensor for the material used and test cycle the system.
•••• Sub-Step (a) ••••
[1] Press the “Label” menu from the Offline Home
Screen.
[2] Press the “Calibrate Gap Sensor” button shown in
the top right image.
[3] Thread a section of liner only (labels removed)
under the label gap sensor and place under
tension, see middle right image.
[4] Press the “Calibrate” button to set the material for
operation. During the calibrate process, the Status
LED will flash green while calibrating.
[5] When calibration is complete, the Status LED will
either be green if the process was successful or red
if it failed. The user interface will also show this
information in text.
•••• Sub-Step (b) ••••
Press the “Feed” button to advance a label. If this is a
pneumatic system, make sure there is not already a
label on the tamp pad prior to pressing the feed button.
Adjust the label dispense position and speed for the
best presentation.
Diagraph - an ITW Company LA/4700 User Manual - Revision B
Setup
Page 30