25
450166 • DEWE3-PA8 • Technical Reference Manual • Printing version 1.0.3 • September 11, 2020
MAIN SYSTEM
TRION3-1810M-POWER-4
135
TRION3-1810M-POWER-4
450124 • TRION™ series modules • Technical Reference Manual • Printi ng version 1.4.1 • September 08, 2020
TRION3™ module for 4-phase power analysis
Channels:
4 power channels
Sampling:
up to 10 MS/s
Resolution:
18-bit
Voltage input:
1000 V
RMS
Modular input:
Current (CAT II): 0.2 A
RMS
to 20 A
RMS
Voltage (not isolated): 5 V
RMS
/ 1 V
RMS
Voltage (CAT II): 600 V
RMS
/ 5 V
RMS
Module specifi cations
TRION3-1810M-POWER-4 with TRION-POWER-SUB
-dL
V-1
TRION3-1810M-POWER-4 with TRION-POWER-SUB
-CUR-xA-1B
continued on next page ...
TRION3-1810M-POWER-4
Input channels
4 Power channels, each with one voltage and one current input
Resolution
18-bit
Sample Rate
up to 10 MS/sec; 512 MByte onboard data buff er
The following are all accuracy conditions referenced in this section:
Temperature: 23±5°C, Humidity: 40 to 60 % RH, Input waveform: sine wave, Common mode voltage: 0 V, Line fi lter: Auto (8th order Butterworth), Sample rate: 10 MS/s,
Resolution: 24 bit, Driver Version >= 3.7, Power factor: 1, After warm-up. After zero level, 12-month Accuracy ±(Reading error + Range error), Accuracy: Frequency (f) in [kHz]
Fixed high-voltage inputs U1 .. U4 (permanently installed)
Input range 1000 V (±2000 V
PEAK
) CF = 2
Accuracy
1) 2)
DC:
±0.02 % of reading ±0.02 % of range
0.5 Hz to 1 kHz:
±0.03 % of reading
1 kHz to 5 kHz:
±0.15 % of reading
5 kHz to 10 kHz:
±0.35 % of reading
10 kHz to 50 kHz:
±0.6 % of reading
50 kHz to 300 kHz:
±(0.02 % * f) of reading
Gain drift 20 ppm / °C
Off set drift
5 mV / °C
Typical THD -95 dB
CMRR >85 dB @ 50 Hz; >60 dB @ 1 kHz; >40 dB @ 100 kHz
Bandwidth 5 MHz
Safety CAT IV 600 V / CAT III 1000 V
Instantaneous maximum allowable input 4000 V
PEAK
or 3000 V
RMS
(1s)
Continuous maximum allowable input 2000 V
RMS
or 2000 V
DC
Input resistance
5 MΩ; 2 pF
Isolation (earth) resistance
100 GΩ; 2.5 pF
Connector Safety banana sockets
1)
add 0.02 % of reading with fi lter settings OFF
2)
below 1 % of range, add 10 ppm of range
Interchangeable sub-modules (I1 .. I4)
20 A module
TRION-POWER-SUB-CUR-20A-1B
Range 20 A (±40 A
PEAK
)
Accuracy
1) 2)
DC:
±0.02 % of reading ±0.02 % of range
3)
0.5 Hz to 1 kHz:
±0.03 % of reading
1 kHz to 5 kHz:
±0.15 % of reading
5 kHz to 10 kHz:
±0.35 % of reading
10 kHz to 50 kHz:
±(0.3 % + 0.05 % * f) of reading
50 kHz to 300 kHz:
±(0.10 % * f) of reading
Safety CAT II 600 V, unfused
Bandwidth 300 kHz
Connector Safety banana sockets (male)
Instantaneous maximum allowable input 50 A
PEAK
or 40 A
RMS
(1s)
Continuous maximum allowable input 20 A
RMS
Input resistance
2 mΩ
1)
For self-generated heat caused by current input, add 0.00008 × I² % of r 15 × I² μA to the current accuracy. ‘I‘ is the current reading [A]. The infl uence from self-
generated heat continues until the temperature of the shunt resistor inside the DEWE3-Chassis lowers even if the current input changes to a small value.
2)
below 1 % of range, add 50 ppm of range
3)
add 0.03 % of range with no zero level