MAIN SYSTEM
Sensor power supply
Compatible sensors/transducers (selection)
Output Power (SUM)
150 W
Output C15 V Line (Single/Sum) 1.5 A / 5 A
Output Current -15 V Line (Single/Sum) 1.5 A / 5 A
Output C9 V Line (Single/Sum)
0.6 A / 0.6 A
Output Current (Single Plug)
1.5 A
PA-IT-60/65
±15 V
PA-IT-200/205
±15 V
PA-IT-400/405
±15 V
PA-IT-700
±15 V
PA-IT-1000
±15 V
PA-IN-1000-S
±15 V
PA-IN-2000-S
±15 V
PA-LF-X
±15 V
SE-CUR-CLAMP-150-DC
±15 V
SE-CUR-CLAMP-200-DC
±15 V
SE-CUR-CLAMP-500-DC
±15 V
SE-CUR-CLAMP-1000-DC
±15 V
SE-CUR-LFR
+9 V
Known limitations PA-IT-1000 and PA-IN-2000-S
Due to the maximum aggregated output current of 5 A, not more than 4x PA-IT-1000 or PA-IN-2000-S can be supplied
in the absolute worst case due to the concurrency factor of the applied power system. The following applications can
be met (selection) with nominal current due to a concurracy factor less than one:
> 4x 1-phase DC systems (4x PA-IT-1000 or PA-IN-2000-S transducers)
> 4x 2-phase DC systems (8x PA-IT-1000 or PA-IN-2000-S transducers)
> 2x 3-phase AC s 1x 1-phase DC system (7x PA-IT-1000 or PA-IN-2000-S transducers)
> 1x 6-phase AC 1 x 1-phase DC system (7x PA-IT-1000 or PA-IN-2000-S transducers)
!
!
3 2 1
4 5 6
Lemo ERA.1S.306
Pin assignment
1: +15 V
2: -15 V
3: +9 V
4: DGND
5: n.c.
6: DGND
NOTE: Do not use the zero-flux
transducer system without
power supply. Induction of
currents can damage the
built-in electronics (see user
manual).
NOTE: The maximum cable lenght
from the transducer to the
device is 5 m. Longer cable
lenghts may cause a too high
voltage drop.