MAIN SYSTEM
1
11.6" multi-touch display
The DEWE3-PA8 is equipped with a bright 11.6" wide screen multi-touch panel (1920 x 1080, full HD) to control the
instrument. Familiar gestures such as pinch and zoom are fully implemented within the operating system and will be
Operating with the touchscreen
'.
2
Power on/off switch
The power on/off switch at the front of the system is used to switch on the system. It only works if the main power
switch
(14)
is switched to position 'I'.
3
LED display for load of current clamps
The 10 Segment LED display indicates the load attached to the power supply connectors for current clamps (
18
) at
the rear of the instrument. One LED segment equals 10 %. When exceeding 80 % during active power measurement,
consider using an additional DW2-CLAMP-DC-POWER-8 box to split the power (e.g. when using 8 * PA-IT-1000).
Further information refer to chapter '
4
USB2.0 interface connectors (Universal Serial Bus, 6x)
The USB2.0 interface connectors meet standard USB pin assignment.
5
USB3.1 interface connectors (Universal Serial Bus, 4x)
The USB3.1 Gen 1 interface connectors meet standard USB pin assignment.
6
Intake vent and filter pad
7
HDD/SSD drive bay for high speed drives
The DEWE3-PA8 comes with a preinstalled 1 TB HDD dedicated for data storage (drive bay at the front).
Optional upgrades:
> SSD-PCIe-1T-2T:
Upgrade from 1 TB to 2 TB industrial grade, PCIe attached solid state disk
8
HDD/SSD drive bay for operating system
Additionally to the high speed drive at the front, the DEWE3-PA8 is equipped with a 120 GB SSD for operating system
and application software (drive bay at the rear panel).
CAUTION:
Information for systems with SSD drives
Wait for 40 seconds after big files were deleted. The HDD activity LED is lit to indicate that the SSD
is deleting the file and TRIM/garbage collection is in progress. Wait until the process is finished
before you start to write the next file.
9
Power supply input connector for redundant SENSOR supply
Input range: 100 .. 240 VAC (power cord included).
For details see chapter '
10 Power switch for redundant SENSOR supply (current clamps)
This switch turns on/off the power supply for the current clamp connectors.
Further information refer to chapter '
'.
!