27
450166 • DEWE3-PA8 • Technical Reference Manual • Printing version 1.0.3 • September 11, 2020
MAIN SYSTEM
137
TRION3-1810M-POWER-4
450124 • TRION™ series modules • Technical Reference Manual • Printi ng version 1.4.1 • September 08, 2020
Clamp input module
TRION-POWER-SUB-dLV-1
Range 5 V (±10 V
PEAK
)
NOT ISOLATED
Accuracy
1)
DC:
±0.02 % of reading ±0.02 % of range
0.5 Hz to 5 kHz:
±0.03 % of reading
5 kHz to 30 kHz:
±(0.01 % * f) of reading
30 kHz to 50 kHz:
±(0.02 % * f) of reading
50 kHz to 100 kHz:
±(0.1 % * f) of reading
Typical THD -95 dB
Typical CMRR 80 dB @ 50 Hz; 80 dB @ 1 kHz; 65 dB @ 10 kHz; 45 dB @ 100 kHz
Safety Depending on connected clamp
Absolute maximum voltage ±30 V
Bandwidth 100 kHz
Connector DSUB-9
Input resistance
1 MΩ
Sensor supply (+9 V) max. 40 mA
1)
below 1 % of range, add 10 ppm of range
Voltage input 5 V (600 V isolated) module
TRION-SUB-5V
Range 5 V (±10 V
PEAK
) CF=2
Accuracy DC:
±0.02 % of reading ±0.005 % of range
0.5 Hz to 1 kHz:
±0.03 % of reading
1 kHz to 100 kHz:
±(0.02 % * f) of reading
100 kHz to 200 kHz:
±(0.04 % * f) of reading
Gain drift 20 ppm / °C
Offset drift
1 µV / °C
Typical THD -102 dB
Typical CMRR >140 dB @ 50 Hz; >106 dB @ 10 kHz; >102 dB @ 100 kHz
Bandwidth (-3dB) 300 kHz
Rated voltage according to EN 61010-2-30 600 V CAT II
Isolation 600 V
RMS
; 35 kV/µs transient immunity
Common mode voltage 600 V
RMS
Transient immunity 35 kV/µs
Overvoltage protection 600 V
RMS
or 1000 V
DC
Input impedance
5 MΩ; 22 pF
Isolation (earth) resistance
100 GΩ; 4 pF (IN- to GND)
Connector Safety banana sockets
Voltage input 600 V CATII module
TRION-SUB-600V
Range 600 V (±1500 V
PEAK
) CF=2.5
Accuracy DC:
±0.02 % of reading ±0.005 % of range
0.5 Hz to 10 kHz:
±0.03 % of reading
10 kHz to 100 kHz:
±(0.015 % * f) of reading
100 kHz to 200 kHz:
±(0.04 % * f) of reading
Gain drift 20 ppm / °C
Offset drift
1 mV / °C
Typical THD -105 dB
typical CMRR
>100 dB @ 50 Hz; >90 dB @ 1 kHz; >70 dB @ 10 kHz; >50 dB @ 100 kHz
Bandwidth (-3dB) 300 kHz
Rated voltage according to EN 61010-2-30 600 V CAT II
Isolation 600 V
RMS
; 35 kV/µs transient immunity
Common mode voltage 600 V
RMS
Transient immunity 35 kV/µs
Overvoltage protection 1500 V
PEAK
(1s)
1000 V
RMS
or 1500 V
DC
Input impedance
5 MΩ; 3.5 pF
Isolation (earth) resistance
100 GΩ; 4 pF (IN- to GND)
Connector Safety banana sockets
continued on next page ...
continued from previous page ...
!
!
Содержание DEWE3-PA8
Страница 4: ...4 PREFACE Notes ...
Страница 6: ...6 TABLE OF CONTENT ...
Страница 44: ...C2 NOTES ...