©2022 by DermLite LLC | 01 Mar 2022 | DL200HY-1701D
(English)
Instructions
Intended Use
This device is intended for medical purposes to illuminate body surfaces. It is
used for non-invasive visual examination of intact skin.
This battery-operated product is designed for external examination only within
professional healthcare facilities by medical professionals.
Check the correct operation of the device before use! Do not use the device if
there are visible signs of damage.
CAUTION: Do not look directly into the LED light. Patients must close their eyes
during examinations.
In case of a serious incident with the use of this device, notify DermLite immedi-
ately and, if required by local regulations, your national health authority.
CAUTION: Do not use the device in fire or explosive risk area (e.g. oxygen-rich
environment).
WARNING: This product can expose you to chemicals including methylene
chloride and hexavalent chromium, which are known to the State of California to
cause cancer or reproductive toxicity. For more information go to www.P65Warn-
ings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC Emissions and Immunity level requirements
of the standard IEC 60601-1-2:2014. The emission characteristics of this equip-
ment make it suitable for use in professional healthcare environment as well as
residential environment (CISPR 11 Class B). This equipment offers adequate
protection to radio communication service. In the rare event of interference to the
radio communication service, the user might need to take mitigation measures,
such as relocating or re-orienting equipment.
WARNING: Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
WARNING: Use of accessories other than those provided by the manufacturer of
this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF transmitters should be used no closer than 30 cm (12
inches) to any part of the device. Otherwise, degradation of the performance of
this equipment could result.
IMPORTANT: Before initial use, please charge the internal lithium ion battery.
See “Battery and Charging” for details.
DermLite DL200 Hybrid is a smartphone-compatible pocket dermatoscope with
10x magnification designed to examine skin lesions with polarized or non-polar
-
ized illumination, with or without skin contact.
IceCap®
To support your infection control measures, snap one of the included disposable
IceCaps (IC) over the installed faceplate (FP). To remove and dispose, simply
pull it off.
Hold your DermLite DL200 so that the LEDs are aimed at the skin. Position the
device ©5mm above the skin and push the Power button (P) to turn on the de-
vice. Look through the lens with your eye 25-50 mm above your DermLite. Move
the device closer or farther away from the skin to obtain the desired image focus.
For dermoscopy with skin contact, or for camera use, turn the Dial (D) to extend
the Spacer (S) and focus the image.
To toggle between non-polarized and polarized illumination, push the Mode
button (M). The Polarization Indicator (PI) lights up in polarized mode.
To turn off your DermLite, push the Power button (P); it turns off automatically
after 3 minutes.
Removing the Spacer
The Spacer (S) may be removed by turning the Dial (D) counter-clockwise be-
yond a noticeable resistance, then pulling it off the device. To attach the Spacer,
insert its 3 legs into the 3 holes until resistance is felt. Place the device level on
a flat surface and evenly push down while simultaneously rotating the Dial (D)
clockwise.
Removing the Faceplate
The Faceplate (FP) may simply be pulled off the Spacer (S) by grabbing it at the
fingernail notches. To reattach, simply snap it back in place.
Photography
Your DermLite DL200 may be attached to many cameras or smartphones using
DermLite MagnetiConnect® adapters (sold separately). Prior to taking photos,
wipe the Faceplate (FP) clean and extend the Spacer (S).
Battery and Charging
CAUTION: This device uses a special 3.7V 1800mAh lithium ion battery, which
can only be purchased from DermLite or from an authorized DermLite dealer. Do
not under any circumstances use another battery other than the one designed for
this unit.
This device is equipped with a four-level Charge Indicator (CI). When the unit is
turned on and fully charged, all four LEDs light up. When less than 75% of battery
life remains, three LEDs are illuminated. If more than half the battery is drained,
two LEDs are on, while a single LED indicates that less than 25% of battery
capacity remains. To charge your device, connect the supplied USB-C to USB
cable to the Charging Port (CP) and any IEC 60950-1 compliant (5V) USB port.
The Charge Indicator (CI) pulses to indicate charging, with the number of pulsing
LEDs corresponding with charge level. When charging is complete, all four indi-
cator LEDs illuminate. After years of use, you may wish to re-place the battery,
which is only available directly from DermLite or an authorized DermLite dealer.
To remove the battery (B), detach the snap-on Cover (C): Pry it open, starting at
the small notch (N), using a screwdriver, then work your way around the Cover to
release it completely. Disconnect the Battery Plug (BP), insert a new battery and
reconnect. Reattach the Cover securely.
Troubleshooting
Please check www.dermlite.com for the most current troubleshooting information.
If your device requires servicing, visit www.dermlite.com/service or contact your
local DermLite dealer.
Care and Maintenance
CAUTION: No modification of this equipment is allowed.
Your device is designed for trouble-free operation. Repairs shall be made only by
qualified service personnel.
Cleaning
The exterior of your device, except the optical parts, may be wiped clean with
isopropyl alcohol (70% vol.) prior to use on a patient. The lens should be treated
as high-quality photographic equipment and should be cleaned with standard lens
cleaning equipment and protected from harmful chemicals. Do not use abrasive
material on any part of the equipment or immerse the device in liquid. Do not
autoclave.
Warranty: 10 years for parts and labor. The battery is warranted for 1 year.
Disposal
This device contains electronics and a lithium-ion battery that must be separated
for disposal and may not be disposed of with general household waste. Please
observe local disposal regulations.
The DermLite DL200 set includes:
DL200 dermatoscope (21 LEDs, 25 mm 10x lens, spacer, faceplate with 10 mm
reticle), silicone cover, lanyard, microfiber case, USB charging cable, 4 IceCaps.
Technical Description
Visit www.dermlite.com/technical or contact your local DermLite dealer.