1. HANDPIECE RATIO SELECTOR:

Sets the Handpiece Ratio Indicator on the
Console LCD Display to read accurately for
different handpiece ratios.
Multiple handpieces with different ratios are
often required to obtain the various speed
ranges used in dental procedures.
The available handpiece ratios and speed  &
torque ranges for each are shown in Chart 2
below.

Chart 2 - Ratio, Speed and, Torque

Push the Handpiece Ratio Selector up or down
arrows until the displayed handpiece ratio
matches the handpiece ratio you are using.
Ratio settings are held in memory even after
unit is turned off.
Auto Stop Reverse (ASR) is not available with
Increaser handpieces.

2. SPEED (RPM) SELECTOR:

Controls the speed of the motor and
handpiece. Push the Speed Up arrow to
increase speed and the Speed Down arrow to
decrease speed. The Speed (RPM) Indicator
on the Console LCD Display shows actual
RPM.  For display accuracy, the displayed
handpiece ratio must exactly match the ratio of
the handpiece being used.  Speed settings are
held in memory even after unit is turned off.

3. TORQUE SELECTOR:

Allows the user to adjust exact torque limits
in Gram•Centimeter increments. Push the
Torque Selector up/down arrows until the
desired torque level is indicated on the
Console LCD Display.  By adjusting the
torque control to reduced limits for each
instrument size and operating the handpiece
at a constant RPM, the risk of instrument
breakage is greatly reduced.

Torque settings are held in memory even
after unit is turned off.

For most endodontic root canal applications,
the included 1/8 reduction contra-angle
handpiece is normally used. Proper-contra
angle maintenance, including cleaning and
lubrication are essential to minimize torque
losses when using the torque reduction set-
tings.

4. FORWARD/REVERSE SELECTOR

(FWD/REV):

Determines the rotation of the motor and
handpiece. A red LED light above this button
will indicate the reverse rotation during
operation. When the Reverse Beeper (BPR)
feature is active, an audible beeper (and the
red LED light) will indicate the reverse
rotation.

5. MAXIMUM TORQUE BUTTON (MAX):

Press and release to override the Endo DTC

®

Torque Control feature.

With MAX Torque activated:

The LED above the Max. Torque Button will
display a steady red light, the Torque
Indicator will display "MAX", and the motor
will operate at full torque. When the Max
function is enabled, the Auto Stop Reverse
(ASR) and torque warning tone features are
automatically disabled.

With MAX Torque disabled:

When the motor reaches the programmed
maximum torque level, the LED above the
Max. Torque Button will blink green. This is
the normal setting for endodontic procedures.

ADVANCED OPERATION AND FUNCTIONS

Note:

Speed and torque ranges are accurate when
handpieces are used with the supplied 30K motor.

Handpiece

Speed

Torque

Ratio

Range (RPM)

Range (g•cm)

5/1

1,500 - 150,000

2-15 (No ASR)

1/1 

TUL or AG

300 - 30,000

4 -125

1/4

75 - 7,500

15 - 500

1/8 

TUL or AG

40 - 3,750

30 - 1,000

1/16 

TUL or AG

20 - 1,800

60 - 2,000

1/64

5 - 460

400 - 7,000

Содержание AEU-25T

Страница 1: ...Endo DTC DIGITAL TORQUE CONTROL DUAL VOLTAGE AEU 25T Electronic Endodontic System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...electric handpiece control system that is used in dentistry to drive endodontic files and instruments CONTRAINDICATIONS None Known WARNINGS Equipment not suitable for use in the presence of a flammabl...

Страница 3: ...se features combine to make the perfect rotary file system for endodontic and other dental procedures Congratulations This system is engineered to provide many years of reliable service Please read th...

Страница 4: ...e AE 4B 30 motor to the receptacle on the lower front panel of the console 4 Attach the E Type handpiece to the motor See Figure 2 5 Attach the supplied AE 7P On Off Foot Switch to the connector on th...

Страница 5: ...e Front Figure 4 Console Back Handpiece Ratio Selector Speed Selector Calibration System Setup PRESET MEMORY LOCK RECALL FACTORY SETUP AUTO STOP REVERSE REVERSE BEEPER TORQUE WARNING TONE ACCESSORY WA...

Страница 6: ...an endodontic rotary file into the handpiece 4 Use the Calibration function to calibrate the Endo DTC See Caution at right Press and release the CAL button The unit will run for a few seconds measuri...

Страница 7: ...sal S2 F1 File 3 ProTaper Universal F2 F3 F4 F5 File 4 ProTaper Universal Retreatment D1 D2 File 5 ProTaper Universal Retreatment D3 Preset Button 4 Default File Series DENTSPLY Tulsa Dental Specialti...

Страница 8: ...strument breakage is greatly reduced Torque settings are held in memory even after unit is turned off For most endodontic root canal applications the included 1 8 reduction contra angle handpiece is n...

Страница 9: ...To exit the System Setup Menu press 3 SAVE CHANGES selection will appear Select 1 to save changes and exit or select 2 to reject all changes and exit 6 CALIBRATION AND SYSTEM SETUP BUTTON Calibration...

Страница 10: ...dible when a GT Accessory file or ITR Emulation Accessory file is in use This is a system wide setting and is not saved on a Preset Button when programming To disable the Accessory Warning press 1 To...

Страница 11: ...e file series shown in Chart 1 on page 5 The Preset Buttons are also used to navigate the various menus in the System Setup and File Library The Preset Buttons can be customized to suit individual pre...

Страница 12: ...you have completed the customiza tion of the desired preset buttons re enable the Preset Memory Lock to avoid inadvertently entering the File Library during a procedure For instructions read about the...

Страница 13: ......

Страница 14: ...d inadvertently entering the File Library dur ing a procedure For instructions read about the System Setup Menu beginning on page 6 1 Disable the Preset Memory Lock For instructions read about the Sys...

Страница 15: ...ge 6 2 Set the Handpiece Ratio to the correct set ting Press the Handpiece Ratio Selector up or down arrows until the Handpiece Ratio Indicator matches the ratio of the handpiece to be used 3 Set the...

Страница 16: ...e from the motor and install plug as shown at right The included motor handpiece holding cradle is also fully autoclavable CONSOLE The exterior of the console may be cleaned by wiping with a soft clot...

Страница 17: ...voltage replace the fuses with fuses that are rated for the new voltage Turn the Fuse holder until the white lettering with the proper voltage is right side up 4 Replace the Fuse holder and close the...

Страница 18: ...5 volts 0 5A 250V slo blo fuse for 230 volts Check motor plug connection Check foot switch connection Depress foot switch Check that a file is properly seated in the handpiece and the latch is closed...

Страница 19: ...anty any risks or liabilities arising from the clinical use of its products whether or not such use involves coincidental utilization of products manufactured by others Tulsa Dental Specialties makes...

Страница 20: ...Johnson City TN 37604 1 800 662 1202 Fax 1 800 597 2779 www TulsaDentalSpecialties com Rx Only DFUDTC Rev 4 12 12 P N 420322 Rev J ECO 12954 12 12 Software P N 890002 Rev J TUL06 Series 2012 DENTSPLY...

Отзывы: