background image

8

REVERSE BEEPER
1=on, 2=off, 3=esc

When the Reverse Beeper (BPR) feature is active,
an audible beep will sound when the handpiece is
rotating in the reverse direction.  This is a system-
wide setting and is not saved on a Preset Button
when programming.

• To enable the Reverse Beeper feature, press 1.

• To disable the Reverse Beeper feature, press 2.

• To exit the System Setup Menu, press 3.

SAVE CHANGES 

selection will appear. Select 1

to save changes and exit or select 2 to reject all
changes and exit.

TORQUE WARNING TONE
1=on, 2=off, 3=esc

When the Torque Warning feature is active, a
chirping warning noise will be audible when the
torque level reaches approximately 75% of the
selected torque limit. This is a system-wide setting
and is not saved on a Preset Button when
programming.

• To enable the Torque Warning Tone,

press 1.

• To disable the Torque Warning Tone,

press 2.

• To exit the System Setup Menu, press 3.

SAVE CHANGES

selection will appear. Select 1

to save changes and exit or select 2 to reject all
changes and exit.

ACCESSORY WARNING TONE

1=off, 2=on, 3=esc

When the Accessory Warning feature is active, a
warning tone will be audible when a GT Accessory
file or ITR Emulation Accessory file is in use. This is
a system-wide setting and is not saved on a Preset
Button when programming.

• To disable the Accessory Warning, press 1.

• To enable the Accessory Warning press 2.

• To exit the System Setup, press 3. 

SAVE CHANGES 

selection will appear.

Select 1 to save changes and exit or select
2 to reject all changes and exit.

SLEEP MODE
1=on, 2=off, 3=esc

After 20 minutes of inactivity, the Endo DTC

®

will enter Sleep Mode to conserve power. In
Sleep Mode, the lighted Console LCD display
will turn off and the green LEDs above the
CAL, FWD/REV, and MAX buttons will blink
once every second.
To resume normal operation when the Endo
DTC

®

is in Sleep Mode, press any button on

the console or press the foot control.
This is a system-wide setting and is not saved
on a Preset Button when programming.

• To enable the Sleep Mode, press 1.

• To disable the Sleep Mode, press 2.

• To exit the System Setup, press 3. 

SAVE

CHANGES

selection will appear. Select 1 to

save changes and exit or select 2 to reject
all changes and exit.

7. POWER SWITCH:

Controls power on/off to console.  On the
switch, the on position is represented by a
‘1’ and the off position is represented by a
‘0’. When the AEU-25 is turned on and not
in Sleep Mode, the Console LCD Display
will light.

SYSTEM SETUP MENU - CONTINUED:

Содержание AEU-25T

Страница 1: ...Endo DTC DIGITAL TORQUE CONTROL DUAL VOLTAGE AEU 25T Electronic Endodontic System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...electric handpiece control system that is used in dentistry to drive endodontic files and instruments CONTRAINDICATIONS None Known WARNINGS Equipment not suitable for use in the presence of a flammabl...

Страница 3: ...se features combine to make the perfect rotary file system for endodontic and other dental procedures Congratulations This system is engineered to provide many years of reliable service Please read th...

Страница 4: ...e AE 4B 30 motor to the receptacle on the lower front panel of the console 4 Attach the E Type handpiece to the motor See Figure 2 5 Attach the supplied AE 7P On Off Foot Switch to the connector on th...

Страница 5: ...e Front Figure 4 Console Back Handpiece Ratio Selector Speed Selector Calibration System Setup PRESET MEMORY LOCK RECALL FACTORY SETUP AUTO STOP REVERSE REVERSE BEEPER TORQUE WARNING TONE ACCESSORY WA...

Страница 6: ...an endodontic rotary file into the handpiece 4 Use the Calibration function to calibrate the Endo DTC See Caution at right Press and release the CAL button The unit will run for a few seconds measuri...

Страница 7: ...sal S2 F1 File 3 ProTaper Universal F2 F3 F4 F5 File 4 ProTaper Universal Retreatment D1 D2 File 5 ProTaper Universal Retreatment D3 Preset Button 4 Default File Series DENTSPLY Tulsa Dental Specialti...

Страница 8: ...strument breakage is greatly reduced Torque settings are held in memory even after unit is turned off For most endodontic root canal applications the included 1 8 reduction contra angle handpiece is n...

Страница 9: ...To exit the System Setup Menu press 3 SAVE CHANGES selection will appear Select 1 to save changes and exit or select 2 to reject all changes and exit 6 CALIBRATION AND SYSTEM SETUP BUTTON Calibration...

Страница 10: ...dible when a GT Accessory file or ITR Emulation Accessory file is in use This is a system wide setting and is not saved on a Preset Button when programming To disable the Accessory Warning press 1 To...

Страница 11: ...e file series shown in Chart 1 on page 5 The Preset Buttons are also used to navigate the various menus in the System Setup and File Library The Preset Buttons can be customized to suit individual pre...

Страница 12: ...you have completed the customiza tion of the desired preset buttons re enable the Preset Memory Lock to avoid inadvertently entering the File Library during a procedure For instructions read about the...

Страница 13: ......

Страница 14: ...d inadvertently entering the File Library dur ing a procedure For instructions read about the System Setup Menu beginning on page 6 1 Disable the Preset Memory Lock For instructions read about the Sys...

Страница 15: ...ge 6 2 Set the Handpiece Ratio to the correct set ting Press the Handpiece Ratio Selector up or down arrows until the Handpiece Ratio Indicator matches the ratio of the handpiece to be used 3 Set the...

Страница 16: ...e from the motor and install plug as shown at right The included motor handpiece holding cradle is also fully autoclavable CONSOLE The exterior of the console may be cleaned by wiping with a soft clot...

Страница 17: ...voltage replace the fuses with fuses that are rated for the new voltage Turn the Fuse holder until the white lettering with the proper voltage is right side up 4 Replace the Fuse holder and close the...

Страница 18: ...5 volts 0 5A 250V slo blo fuse for 230 volts Check motor plug connection Check foot switch connection Depress foot switch Check that a file is properly seated in the handpiece and the latch is closed...

Страница 19: ...anty any risks or liabilities arising from the clinical use of its products whether or not such use involves coincidental utilization of products manufactured by others Tulsa Dental Specialties makes...

Страница 20: ...Johnson City TN 37604 1 800 662 1202 Fax 1 800 597 2779 www TulsaDentalSpecialties com Rx Only DFUDTC Rev 4 12 12 P N 420322 Rev J ECO 12954 12 12 Software P N 890002 Rev J TUL06 Series 2012 DENTSPLY...