RECALL FACTORY SETUP
1=no, 2=yes, 3=esc
Recalling the Factory Setup allows you to reset
your Endo DTC
®
to factory specifications. NOTE:
Any customized presets will be lost if you Recall
Factory Setup.
• To keep your customized settings, press 1.
• To return the Endo DTC
®
preset buttons to the
factory settings (as shown in Chart 1 on page
5), press 2.
•
CONFIRM RECALL
will appear. Select 2 to
confirm recall or select 1 to reject changes.
• To exit the System Setup Menu and save
changes, press 3.
AUTO STOP REVERSE
1=on, 2=off, 3=esc
When the Auto Stop Reverse (ASR) feature is
enabled, rotation of the handpiece will automati-
cally stop and reverse when the selected torque
limit is reached. Do not release the foot pedal
when this occurs; keep the foot pedal activated
and the Endo DTC
®
will automatically alternate
between forward and reverse rotation in an
attempt to free the instrument.
The Auto Stop Reverse setting is saved when
programming Preset Buttons.
• To enable the Auto Stop Reverse feature,
press 1.
This setting is
highly recommended
for
endodontic procedures.
• To disable the Auto Stop Reverse feature,
press 2.
Disabling auto stop reverse is
not recom-
mended
for endodontic procedures, as file
separation may occur.
• To exit the System Setup Menu, press 3.
SAVE CHANGES
selection will appear.
Select 1 to save changes and exit or select 2
to reject all changes and exit.
6. CALIBRATION AND
SYSTEM SETUP BUTTON:
Calibration (CAL):
Press and release the "CAL" button to activate
the Integrated Calibration System. The AEU-25
will automatically run the handpiece and
system through a self test.
Use the Calibration function in the following
circumstances:
• At the beginning of each day
• After changing handpieces
(especially when changing handpiece type)
• After changing files or burs (optional)
System Setup Menu:
Press and hold the "CAL" button until you hear
a double beep to enter the System Setup Menu.
After entering the System Setup Menu, the fol-
lowing setup options will be displayed on the
Console LCD Display.
PRESET MEMORY LOCK
1=on, 2=off, 3=esc
By default, the Preset Memory Lock is activated
to prevent the user from inadvertently entering
the File Library during a procedure.
To customize the Preset Buttons with other File
Library Settings or User Defined Presets, the
Preset Memory Lock must be disabled.
To protect any customized presets, re-enable
the Preset Memory Lock after programming the
Preset Buttons.
This is a system-wide setting and is not saved
on a Preset Button when programming.
• To enable the Preset Memory Lock,
press 1.
• To disable the Preset Memory Lock,
press 2.
• To exit the System Setup Menu without
changes, press 3.