background image

7

RECALL FACTORY SETUP

1=no, 2=yes, 3=esc

Recalling the Factory Setup allows you to reset
your Endo DTC

®

to factory specifications. NOTE:

Any customized presets will be lost if you Recall
Factory Setup.

• To keep your customized settings, press 1.

• To return the Endo DTC

®

preset buttons to the

factory settings (as shown in Chart 1 on page
5), press 2.

CONFIRM RECALL

will appear. Select 2 to

confirm recall or select 1 to reject changes.

• To exit the System Setup Menu and save

changes, press 3.

AUTO STOP REVERSE
1=on, 2=off, 3=esc

When the Auto Stop Reverse (ASR) feature is
enabled, rotation of the handpiece will automati-
cally stop and reverse when the selected torque
limit is reached.  Do not release the foot pedal
when this occurs; keep the foot pedal activated
and the Endo DTC

®

will automatically alternate

between forward and reverse rotation in an
attempt to free the instrument.

The Auto Stop Reverse setting is saved when
programming Preset Buttons.

• To enable the Auto Stop Reverse feature,

press 1.
This setting is 

highly recommended

for

endodontic procedures.

• To disable the Auto Stop Reverse feature,

press 2.
Disabling auto stop reverse is 

not recom-

mended

for endodontic procedures, as file

separation may occur.

• To exit the System Setup Menu, press 3.

SAVE CHANGES 

selection will appear.

Select 1 to save changes and exit or select 2
to reject all changes and exit.

6. CALIBRATION AND

SYSTEM SETUP BUTTON:

Calibration (CAL): 

Press and release the "CAL" button to activate
the Integrated Calibration System. The AEU-25
will automatically run the handpiece and
system through a self test.

Use the Calibration function in the following
circumstances:

• At the beginning of each day

• After changing handpieces

(especially when changing handpiece type)

• After changing files or burs (optional)

System Setup Menu:

Press and hold the "CAL" button until you hear
a double beep to enter the System Setup Menu.
After entering the System Setup Menu, the fol-
lowing setup options will be displayed on the
Console LCD Display.

PRESET MEMORY LOCK
1=on, 2=off, 3=esc

By default, the Preset Memory Lock is activated
to prevent the user from inadvertently entering
the File Library during a procedure.
To customize the Preset Buttons with other File
Library Settings or User Defined Presets, the
Preset Memory Lock must be disabled.
To protect any customized presets, re-enable
the Preset Memory Lock after programming the
Preset Buttons.
This is a system-wide setting and is not saved
on a Preset Button when programming.

• To enable the Preset Memory Lock,

press 1.

• To disable the Preset Memory Lock,

press 2.

• To exit the System Setup Menu without

changes, press 3. 

Содержание AEU-25T

Страница 1: ...Endo DTC DIGITAL TORQUE CONTROL DUAL VOLTAGE AEU 25T Electronic Endodontic System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Страница 2: ...electric handpiece control system that is used in dentistry to drive endodontic files and instruments CONTRAINDICATIONS None Known WARNINGS Equipment not suitable for use in the presence of a flammabl...

Страница 3: ...se features combine to make the perfect rotary file system for endodontic and other dental procedures Congratulations This system is engineered to provide many years of reliable service Please read th...

Страница 4: ...e AE 4B 30 motor to the receptacle on the lower front panel of the console 4 Attach the E Type handpiece to the motor See Figure 2 5 Attach the supplied AE 7P On Off Foot Switch to the connector on th...

Страница 5: ...e Front Figure 4 Console Back Handpiece Ratio Selector Speed Selector Calibration System Setup PRESET MEMORY LOCK RECALL FACTORY SETUP AUTO STOP REVERSE REVERSE BEEPER TORQUE WARNING TONE ACCESSORY WA...

Страница 6: ...an endodontic rotary file into the handpiece 4 Use the Calibration function to calibrate the Endo DTC See Caution at right Press and release the CAL button The unit will run for a few seconds measuri...

Страница 7: ...sal S2 F1 File 3 ProTaper Universal F2 F3 F4 F5 File 4 ProTaper Universal Retreatment D1 D2 File 5 ProTaper Universal Retreatment D3 Preset Button 4 Default File Series DENTSPLY Tulsa Dental Specialti...

Страница 8: ...strument breakage is greatly reduced Torque settings are held in memory even after unit is turned off For most endodontic root canal applications the included 1 8 reduction contra angle handpiece is n...

Страница 9: ...To exit the System Setup Menu press 3 SAVE CHANGES selection will appear Select 1 to save changes and exit or select 2 to reject all changes and exit 6 CALIBRATION AND SYSTEM SETUP BUTTON Calibration...

Страница 10: ...dible when a GT Accessory file or ITR Emulation Accessory file is in use This is a system wide setting and is not saved on a Preset Button when programming To disable the Accessory Warning press 1 To...

Страница 11: ...e file series shown in Chart 1 on page 5 The Preset Buttons are also used to navigate the various menus in the System Setup and File Library The Preset Buttons can be customized to suit individual pre...

Страница 12: ...you have completed the customiza tion of the desired preset buttons re enable the Preset Memory Lock to avoid inadvertently entering the File Library during a procedure For instructions read about the...

Страница 13: ......

Страница 14: ...d inadvertently entering the File Library dur ing a procedure For instructions read about the System Setup Menu beginning on page 6 1 Disable the Preset Memory Lock For instructions read about the Sys...

Страница 15: ...ge 6 2 Set the Handpiece Ratio to the correct set ting Press the Handpiece Ratio Selector up or down arrows until the Handpiece Ratio Indicator matches the ratio of the handpiece to be used 3 Set the...

Страница 16: ...e from the motor and install plug as shown at right The included motor handpiece holding cradle is also fully autoclavable CONSOLE The exterior of the console may be cleaned by wiping with a soft clot...

Страница 17: ...voltage replace the fuses with fuses that are rated for the new voltage Turn the Fuse holder until the white lettering with the proper voltage is right side up 4 Replace the Fuse holder and close the...

Страница 18: ...5 volts 0 5A 250V slo blo fuse for 230 volts Check motor plug connection Check foot switch connection Depress foot switch Check that a file is properly seated in the handpiece and the latch is closed...

Страница 19: ...anty any risks or liabilities arising from the clinical use of its products whether or not such use involves coincidental utilization of products manufactured by others Tulsa Dental Specialties makes...

Страница 20: ...Johnson City TN 37604 1 800 662 1202 Fax 1 800 597 2779 www TulsaDentalSpecialties com Rx Only DFUDTC Rev 4 12 12 P N 420322 Rev J ECO 12954 12 12 Software P N 890002 Rev J TUL06 Series 2012 DENTSPLY...

Отзывы: