12
Example 2:
To go to the F2 and F3 files in the
Protaper series, press Button #1 twice.
The display will show the following:
PROTAPER F2 F3 F4 F5
1=next, 2=ok, 3=esc
8. Press Button #2 to select the displayed files.
9. The display will show:
USE DEFAULT HP?
1=yes, 2=no, 3=esc
NOTE: The default handpiece is the
handpiece that was shipped with the Endo
DTC
®
, specified in Quick Start Operation
Section 2.
a. If you are using the default handpiece
with this file series, press Button #1 and
go on to Step 10.
b. If you are using a handpiece other than
the default handpiece, press Button #2
and choose the correct handpiece. For
instructions, refer to i and ii on page 10.
10. The Preset Button is now set with the
selected single file settings.
For Example 2 above:
The display will now show:
SINGLE FILE PRESET
PROTAPER F2 F3 F4 F5
11. When you have completed the program-
ming of the desired preset buttons, re-
enable the Preset Memory Lock to avoid
inadvertently entering the File Library dur-
ing a procedure. For instructions, read
about the System Setup Menu beginning
on page 6.
1. Disable the Preset Memory Lock.
For instructions, read about the System
Setup Menu beginning on page 6.
2. Press and hold the desired Preset Button for
1 second (until you hear a double beep).
3. The display will show:
ACCESS FILE LIBRARY?
1=yes, 2=no, 3=esc
Press Button #1 to access the File Library.
4. The display will show:
PATHFILE ALL
1=next, 2=ok, 3=esc
Press Button #1 repeatedly until the desired
file series is shown on the display. See
Chart 3 on page 11 for a list of available file
series.
Example 2:
To go to the Protaper
®
Universal series,
press preset button #1 three times.
The display will then show:
PROTAPER UNIVERSAL
1=next, 2=ok, 3=esc
5. Press Button #2 to select the displayed file
series.
6. The display will show:
LOAD ENTIRE SERIES?
1=yes, 2=no, 3=esc
Press Button #2. Do not load entire series.
7. Press Button #1 repeatedly until the desired
file is shown on the display.
See Chart 3 on page 11 for a list of available
files in each file series.
10. PROGRAMMING A PRESET BUTTON WITH A SINGLE FILE
Содержание AEU-25T
Страница 13: ......