Profi-Line

pH electrode for aquaria

For continuous measurement of the pH value in fresh water and sea water aquaria

- Instructions for use: please read carefully and keep in a safe place. -

GB

Congratulations on buying this high-quality pH electrode from DENNERLE. If used properly and treated with due care, this system will provide reliable and accu-
rate readings of the pH value in your aquarium – an essential basis for optimum fish and plant care.
DENNERLE wishes you lasting enjoyment from your aquarium!

Safety instructions

• All cables leading out of the aquarium must be provided with a drip loop to prevent

any water which may run down the cables from reaching items of equipment or the
mains socket outlet.

• Always disconnect all devices located in the water from the power supply before pla-

cing your hands in the aquarium water.

1 Which equipment can the

pH electrode be used with?

The DENNERLE pH electrode for aquaria works perfectly in combination
with with all DENNERLE pH controllers.Equally, it can also be used with
all other standard pH measuring and control devices with BNC connec-
tor. Please observe the instructions for use of the device concerned!

2 What‘s what

BNC connector

Connecting cable, 3 m

Electrode cap

Plastic stem

Electrode tip with membrane

Quali-Safe electrode protector
(contains KCl solution,
3 molar)

3 Connecting

the pH electrode

• Unscrew the lid of the Quali-Safe

electrode protector and carefully
remove the electrode. Note: Any
white salt crust which may appear on
the electrode shaft consists of calci-
um chloride (KCL) and will not affect
measuring performance.

• Rinse electrode tip briefly under run-

ning tap water, in order to remove

residual KCl solution/ the salt crust.

• Connect BNC connector to pH control-

ler or pH measuring device.

4 Calibrating the pH electrode

4.1 Purpose of calibration and calibration interval

Any measuring instrument can only be as accurate as the connected sen-
sor. The pH electrode is thus a vitally important part of your pH measu-
ring device.
As every pH electrode may drift somewhat in the course of time, it is
crucial to calibrate the pH electrode
-

when putting it into operation for the 1st time

-

after the 1st week

-

and then every 4 to 6 weeks (under normal operating conditions).

4.2 Preparations for the calibrati-
on process

Ideal temperature of the buffer solutions:
20-25 °C.
Pour pH 7 and pH 4 buffer solution into two
separate, clean receptacles up to a level of
approx.

2 cm. Pour a little distilled water

into a third receptacle.

TIP: The DENNERLE calibrating station is particularly convenient and saves
valuable buffer solution.
Please note: The accuracy of the calibration depends above all on the qua-
lity of the employed buffer solutions. We thus recommend original DEN-
NERLE buffer solutions. Do not carry out calibration in the supply bottles -
water on the pH electrode may alter the buffer solutions! Dispose of used
buffer solution down the drain.

1

2

3

6

4

GA pH-Elektrode:GA pH-Elektrode

03.11.2008  16:31 Uhr  Seite 6

Содержание Profi-Line CO2 pH electrode

Страница 1: ...i Set pour Electrode pH pour aquariums 4 Operating instructions for pH Elektrode for aquaria 6 Istruzioni per l uso di Sonda pH per acquari 8 Gebruiksanwijzing voor Aquaria pH elektrode 10 D F GB I NL...

Страница 2: ...Sicherheitshinweise Alle Kabel die aus dem Aquarium herausf hren m ssen mit einer Tropfschlaufe ver sehen sein so dass eventuell an ihnen herunterlaufendes Wasser nicht in die Ger te bzw die Netzstec...

Страница 3: ...ektrode und Wasserstand kontrollieren 4 3 Kalibrierung durchf hren pH Elektrode kurz in destilliertem Wasser absp len und in Pufferl sung pH 7 stellen Ca 1 Minute warten bis sich der ange zeigte Messw...

Страница 4: ...tion 4 2 Pr paration du calibrage Temp rature optimale des solutions tampon 20 25 C Remplir deux r cipients s par s et propres d env 2 cm de solution tampon pH 7 et pH 4 chacun En remplir un troisi me...

Страница 5: ...rassement par les algues Ne jamais poser le c ble de l lectrode avec d autres c bles lectri ques sur de longues distances Ainsi les autres circuits lectriques ne risquent pas d influencer les r sultat...

Страница 6: ...electrode protector and carefully remove the electrode Note Any white salt crust which may appear on the electrode shaft consists of calci um chloride KCL and will not affect measuring performance Ri...

Страница 7: ...h light shield to prevent algae formation Fault Electrode responds very sluggishly or no calibra tion possible pH display does not res pond or is unstable Cause Electrode dirty Air bubble in electrode...

Страница 8: ...ui collegato La sonda pH quindi un componente di importanza decisi va del vostro strumento di misurazione del pH Poich ogni sonda pH con il passare del tempo pu stararsi per il fun zionamento dello st...

Страница 9: ...contro la formazione di alghe Non effettuare la posa del cavo della sonda con altri cavi di ali mentazione per lunghi tratti In questo modo gli altri circuiti elettri ci non influenzeranno i risultat...

Страница 10: ...is daarom een component van doorslaggevende bete kenis van uw pH meter Omdat de afstelling van elke pH elektrode in de loop van de tijd iets veranderen kan is het voor het functioneren van de pH meter...

Страница 11: ...mogelijk pH weergave reageert niet of is instabiel Oorzaak Elektrode vervuild Luchtbel in de tap van de elektrode Bufferoplossingen be dorven te oud De elektrode is ver bruikt of defect pH meetapparaa...

Страница 12: ...1 1108 Vertrieb DENNERLE GmbH D 66957 Vinningen Kundenservice DENNERLE GmbH Industriestrasse 4 D 66981 M nchweiler www dennerle de Made in Germany GA pH Elektrode GA pH Elektrode 03 11 2008 16 31 Uhr...

Отзывы: