Lamp Replacement Instructions

 

WARNING: LAMPS ARE FILLED TO VERY HIGH PRESSURES, OFTEN GREATER THAN 200 PSI AT ROOM 
TEMPERATURE. EYE PROTECTION MUST BE WORN. EXTREME CARE MUST ALWAYS BE EXERCISED  
WHEN HANDLING LAMPS. EXPLOSION HAZARDS ARE POSSIBLE (MOST HAPPEN DURING OPERATION,  
WHEN THE PRESSURE IS HIGHER). INTERNAL LAMP PRESSURES CAN DOUBLE WHEN THE LIGHT IS IN  
USE AND GREATLY EXCEED THAT IF IMPROPERLY COOLED. ONLY A QUALIFIED SERVICE TECHNICIAN  
MAY REPLACE THE SYSTEM LAMP. NEVER POINT THE WINDOW OF THE LAMP TO UNPROTECTED TISSUE.

Symptoms indicating need for replacement are:
1.  Low intensity on system light meter.
2.  Lamp fails while in use or does not turn on.

To Remove the Old Lamp
1.  Turn unit off and unplug the power cord from its power source.
2.  Wait for the unit and the lamp to cool. 
WARNING: Lamp internal pressure is much higher when hot. Improper cooling increases risk of explosion.
3.  Unscrew thumbscrews to remove the lamp access plate.
4.  Grasp lamp and slide towards you to remove from light source. WARNING: Do not drop - Lamp can explode!
5.  Unplug the lamp from the 2-pin lamp socket.

To Insert New Lamp
1.  Orient the lamp facing forward in the lamp guides on the chassis.
2.  Push toward the top of the unit until the lamp slides securely onto the lamp pins.
3.  Replace the lamp access plate and tighten thumbscrews.
NOTE: Make sure the access plate fits correctly in the cutout provided. All corners should be even with the 
surrounding surfaces. If the access plate is misaligned, the unit may not receive AC power.

To Dispose of Old Lamp (Steps 1 and 2 must be performed by qualified service technicians) CeraLux lamps 
contain no hazardous materials such as mercury or rare earth elements. Lamps may be disposed of safely once the 
internal pressure of xenon is released. 
To release internal pressure, do the following:
1. Wear safety glasses.
2. Firmly grasp the lamp and cut or twist the pinch off seal at the rear of the lamp. The gas will be heard escaping.
3. Afterwards, the lamp may be disposed in normal trash.

If unsure about the instructions, DO NOT PROCEED! Contact Den-Mat Holdings, LLC at 1-800-445-0345. 
DenMat can also give instructions regarding shipping the lamp back to DenMat and will dispose of the lamp 
properly.

Unscrew thumbscrews  
  to remove access door.

HANDLES

BRASS SOCKETS

19

Содержание Sapphire plus

Страница 1: ...Dental Curing Light with Optional Tooth Whitening Conforms to AAMI Std ES60601 1 IEC Stds60601 1 6 80601 2 60 62366 Certified to CSA Std C22 2 NO 60601 1 6 Instructions for Use...

Страница 2: ...nel 7 Accessories 7 Set up and Operation Instructions 8 Light Meter Radiometer 9 Curing Mode 9 10 Curing Instructions 10 Bleach Whitening Mode 11 12 Whitening Instructions 12 15 Sanitizing and Sterili...

Страница 3: ...placement must be made by a factory authorized representative Lamp replacement and fuse replacement must be done by a qualified service technician or by factory authorized representative Do not touch...

Страница 4: ...e or tip Do not autoclave the light shield provided with the unit Do not autoclave the light guide Removable curing tip is autoclavable Do not use abrasives or liquid chemicals to clean the tip or the...

Страница 5: ...erfere with other medical devices Since there is no guarantee in a particular installation that other device that interference will not occur test the installation prior to use Ensure that other devic...

Страница 6: ...OR SINGLE TOOTH WHITENING IF THE LIGHT OUTPUT IS GREATER THAN 3000 mW cm2 Call DenMat and return the instrument for repair 4 Do not saturate the controller keypad area with disinfectant or other liqui...

Страница 7: ...sor recognizes that they have been inserted and that the xenon arc lamp can be used when needed Control panels for curing bleaching are present The user interfaces offer a wide range of timing increme...

Страница 8: ...ct button B Timer display C Activator button Each handpiece has an activator button to activate the lamp 6 Removable Autoclavable Tip Optical device for exact placement of light 7 Lamp Access Plate Ac...

Страница 9: ...information Light Guide and Tips The Light Guide provided with the Sapphire Plus provides an optical path for transmitting light energy from the source to the tips Both the Light Guide and removable t...

Страница 10: ...ow key to advance directly to the SCOPE MODE or press the RIGHT arrow key to step through the BLEACH SCOPE and CURING MODES and press the SELECT button to enter that mode Warning Scope Mode is no long...

Страница 11: ...be set from 1 to 10 seconds It is recommended that multiple 5 second exposures be used for resins that need a little more time WARNING Single exposures longer than 5 seconds must be limited to use wit...

Страница 12: ...light cable and any curing tips which will be used Ensure they are in good working order Check to make sure there are no burrs or sharp edges or features which could cut a patient or the user of the...

Страница 13: ...dicated Warnings and Precautions start the system After the timing cycle is completed and the light turns off test the bottom surface for hardness using a dental probe If the material is not completel...

Страница 14: ...n of this user s guide Start the lamp by pressing the ACTIVATOR button on the handpiece or the START STOP button on the control panel Stop the lamp by pressing the ACTIVATOR button or START STOP again...

Страница 15: ...dish Use the brush applicator to apply a 1 1 5 mm layer onto the labial surface of the teeth 9 Install a sanitized Sapphire Whitening Crystal in place of the curing tip on the end of the Sapphire Plu...

Страница 16: ...or any cracking or lifting before re applying the whitening gel If the seal has broken from the tissue it is recommended to remove and replace the Paint On Dental Dam and begin with Step 4 17 At the e...

Страница 17: ...mile White In Office 25 with Take Home 22 Combo WHT0030 LumiBrite Chairside Whitening System 035359155 LumiBrite Desensitizing Gel 034159555 Paint On Dental Dam Kit Blue 404210 Paint On Dental Dam Syr...

Страница 18: ...lus System WARNING Unplug the power cord prior to cleaning with any liquid Allow the system to cool prior to cleaning Avoid spilling any liquid on the system Do not use flammable liquids to clean the...

Страница 19: ...he way Rotate the light guide 1 4 to 1 2 turn in the light guide receptacle Look for the display to light up on the handpiece If the display on the handpiece responds try starting the light again with...

Страница 20: ...de 5 Unplug the lamp from the 2 pin lamp socket To Insert New Lamp 1 Orient the lamp facing forward in the lamp guides on the chassis 2 Push toward the top of the unit until the lamp slides securely o...

Страница 21: ...pair at 1 800 433 6628 Transportation When transporting the Sapphire Plus remove the power cord and the light guide from the Sapphire Plus light source Hold the handle or carry it with both hands one...

Страница 22: ...will be evaluated for the cause and extent of failure or damage by Den Mat Holdings LLC Service Representatives Subject to the results of this evaluation Den Mat Holdings LLC will authorize warranty r...

Страница 23: ...posite veneers Core Paste XP Syringe 030654110 The recognized standard in core build up materials in a fast auto mixing syringe that can be easily syringed directly into canals Infinity Syringeable Va...

Страница 24: ...dings LLC All rights reserved 823057820 R 3 07 18CM EC REP MDSS GmbH Schiffgraben 41 30175 Hannover Germany Symbols Glossary available at www denmat com symbols Den Mat Holdings LLC 1017 W Central Ave...

Отзывы: