Troubleshooting

Problem 
The control panel and handpiece display does not light up. Buttons do not respond.

Corrective action
Check the AC electrical outlet, power cord, and power input on the light to verify that all connections are secure 
and that AC power is on at the outlet. Verify that the power switch is in the ON ( I ) position. 

Problem
The control panel display lights up but the handpiece display does not light. Lamp does not start when the 
ACTIVATOR button on the handpiece is pressed. 

Corrective action
Being careful not to look directly at the distal end of the light guide, press the START/STOP button on the 
front panel. If the lamp does not come on, move to the next step. If the lamp comes on, recheck the light guide 
connections. Disconnect the light guide and firmly reinstall, making sure that it is plugged in all the way. Rotate 
the light guide 1/4 to 1/2 turn in the light guide receptacle. Look for the display to light up on the handpiece. If 
the display on the handpiece responds, try starting the light again with the ACTIVATOR button.

Problem
The control panel display lights up and the CONTROL WHEEL and TIMER SELECT button work, but the 
START/STOP button and ACTIVATOR button do not start the lamp.

Corrective action
Contact Den-Mat Holdings, LLC for technical support at 1-800-433-6628.

Problem 
Light output is low or curing of the composite is slow.

Corrective action
Verify the light guide is plugged all the way into the receptacle on the front of the unit. Check the tip for 
contamination, scratches, or breaks. Clean the tip with composite solvent. Check the light guide for damage. 
Follow instructions for checking and replacing the lamp.

If corrective action does not resolve the problem, contact Den-Mat Holdings, LLC for technical support at  
1-800-433-6628. This includes possible lamp replacement or return of the system for repair.

18

Содержание Sapphire plus

Страница 1: ...Dental Curing Light with Optional Tooth Whitening Conforms to AAMI Std ES60601 1 IEC Stds60601 1 6 80601 2 60 62366 Certified to CSA Std C22 2 NO 60601 1 6 Instructions for Use...

Страница 2: ...nel 7 Accessories 7 Set up and Operation Instructions 8 Light Meter Radiometer 9 Curing Mode 9 10 Curing Instructions 10 Bleach Whitening Mode 11 12 Whitening Instructions 12 15 Sanitizing and Sterili...

Страница 3: ...placement must be made by a factory authorized representative Lamp replacement and fuse replacement must be done by a qualified service technician or by factory authorized representative Do not touch...

Страница 4: ...e or tip Do not autoclave the light shield provided with the unit Do not autoclave the light guide Removable curing tip is autoclavable Do not use abrasives or liquid chemicals to clean the tip or the...

Страница 5: ...erfere with other medical devices Since there is no guarantee in a particular installation that other device that interference will not occur test the installation prior to use Ensure that other devic...

Страница 6: ...OR SINGLE TOOTH WHITENING IF THE LIGHT OUTPUT IS GREATER THAN 3000 mW cm2 Call DenMat and return the instrument for repair 4 Do not saturate the controller keypad area with disinfectant or other liqui...

Страница 7: ...sor recognizes that they have been inserted and that the xenon arc lamp can be used when needed Control panels for curing bleaching are present The user interfaces offer a wide range of timing increme...

Страница 8: ...ct button B Timer display C Activator button Each handpiece has an activator button to activate the lamp 6 Removable Autoclavable Tip Optical device for exact placement of light 7 Lamp Access Plate Ac...

Страница 9: ...information Light Guide and Tips The Light Guide provided with the Sapphire Plus provides an optical path for transmitting light energy from the source to the tips Both the Light Guide and removable t...

Страница 10: ...ow key to advance directly to the SCOPE MODE or press the RIGHT arrow key to step through the BLEACH SCOPE and CURING MODES and press the SELECT button to enter that mode Warning Scope Mode is no long...

Страница 11: ...be set from 1 to 10 seconds It is recommended that multiple 5 second exposures be used for resins that need a little more time WARNING Single exposures longer than 5 seconds must be limited to use wit...

Страница 12: ...light cable and any curing tips which will be used Ensure they are in good working order Check to make sure there are no burrs or sharp edges or features which could cut a patient or the user of the...

Страница 13: ...dicated Warnings and Precautions start the system After the timing cycle is completed and the light turns off test the bottom surface for hardness using a dental probe If the material is not completel...

Страница 14: ...n of this user s guide Start the lamp by pressing the ACTIVATOR button on the handpiece or the START STOP button on the control panel Stop the lamp by pressing the ACTIVATOR button or START STOP again...

Страница 15: ...dish Use the brush applicator to apply a 1 1 5 mm layer onto the labial surface of the teeth 9 Install a sanitized Sapphire Whitening Crystal in place of the curing tip on the end of the Sapphire Plu...

Страница 16: ...or any cracking or lifting before re applying the whitening gel If the seal has broken from the tissue it is recommended to remove and replace the Paint On Dental Dam and begin with Step 4 17 At the e...

Страница 17: ...mile White In Office 25 with Take Home 22 Combo WHT0030 LumiBrite Chairside Whitening System 035359155 LumiBrite Desensitizing Gel 034159555 Paint On Dental Dam Kit Blue 404210 Paint On Dental Dam Syr...

Страница 18: ...lus System WARNING Unplug the power cord prior to cleaning with any liquid Allow the system to cool prior to cleaning Avoid spilling any liquid on the system Do not use flammable liquids to clean the...

Страница 19: ...he way Rotate the light guide 1 4 to 1 2 turn in the light guide receptacle Look for the display to light up on the handpiece If the display on the handpiece responds try starting the light again with...

Страница 20: ...de 5 Unplug the lamp from the 2 pin lamp socket To Insert New Lamp 1 Orient the lamp facing forward in the lamp guides on the chassis 2 Push toward the top of the unit until the lamp slides securely o...

Страница 21: ...pair at 1 800 433 6628 Transportation When transporting the Sapphire Plus remove the power cord and the light guide from the Sapphire Plus light source Hold the handle or carry it with both hands one...

Страница 22: ...will be evaluated for the cause and extent of failure or damage by Den Mat Holdings LLC Service Representatives Subject to the results of this evaluation Den Mat Holdings LLC will authorize warranty r...

Страница 23: ...posite veneers Core Paste XP Syringe 030654110 The recognized standard in core build up materials in a fast auto mixing syringe that can be easily syringed directly into canals Infinity Syringeable Va...

Страница 24: ...dings LLC All rights reserved 823057820 R 3 07 18CM EC REP MDSS GmbH Schiffgraben 41 30175 Hannover Germany Symbols Glossary available at www denmat com symbols Den Mat Holdings LLC 1017 W Central Ave...

Отзывы: