8
Sanitizing and Sterilizing
Sanitizing the Whitening Crystal
Clean with a mild, non-abrasive, liquid soap and thoroughly dry all surfaces. Sanitize with solutions containing
glutaraldehyde, low concentrations of hydrogen peroxide, 70% isopropyl alcohol (70% IPA) or other disinfectants
containing mostly inert ingredients.
Sterilizing the Curing Tip
Clean the tip regularly with composite solvent. The removable tip must be sterilized using standard autoclave
procedures for stainless steel instruments (135º C). Use a sterilization bag and leave sealed for future use.
Failure to sterilize tips between patients increases the risk of patient-to-patient infection.
CAUTION: DO NOT AUTOCLAVE THE LIGHT GUIDE!
Sanitizing The Light Guide and Handle
The light guide cannot be autoclaved. It should be sanitized by using a non-abrasive disinfectant. Do not use
glutaraldehyde-type disinfectants.
Cleaning The Sapphire Plus System
WARNING: Unplug the power cord prior to cleaning with any liquid. Allow the system to cool prior to cleaning.
• Avoid spilling any liquid on the system. Do not use flammable liquids to clean the main system or light guide.
If the internal electronics are exposed to any liquid, disconnect the power. Do not attempt to start the system.
Contact the factory for instructions at 1-800-433-6628.
• Clean the main system using only mild soap and water solution as needed. Do not saturate the control panel
with liquid. Wipe with a damp towel and dry completely.
Replacement Parts
Additional tips, shields, curing rings, light guides and lamp modules are available from Den-Mat Holdings, LLC.
See back cover for ordering information.
Service
Minor problems can be quickly identified and resolved. These are usually related to the electrical power
connection or light guide connection. All other malfunctions must be diagnosed and repaired by Den-Mat
Holdings, LLC. Contact Den-Mat Holdings, LLC for technical support at 1-800-433-6628.
WARNING: THE SUPPLIED ORANGE SAFETY GLASSES MUST BE WORN DURING USE OR TESTING.
17