Set-up and Operation Instructions

 

Compatibility of Existing Sapphire

®

 and Sapphire Plus Lights and Devices 

The labeling on the curing handpiece and Sapphire light source will indicate whether these are different than 
the new Sapphire Plus. The curing handpiece and Sapphire Light source may be labeled Sapphire, Rembrandt

®

 

Sapphire, or Sapphire Supreme.
The curing handpieces for the new Sapphire Plus light have been designed to work with existing Sapphire lights 
(reverse compatible). At the same time, curing handpieces for existing Sapphire lights can be used with the new 
Sapphire Plus light (forward compatible).

Using Existing Sapphire Curing Handpieces
When using existing curing devices in the Sapphire Plus, it is necessary to manually select the correct mode of 
operation.
When the Sapphire Plus light is powered up without a device installed in the light guide receptacle on the front 
of the light source, the error message “No HP” (no handpiece) will appear on the display.
When existing curing handpieces are installed in the light source, the display windows will change from “No HP” 
to a flashing “Ch HP” (choose handpiece). This means that a handpiece has been recognized and the operator 
is being prompted to select the correct user mode. Press the LEFT arrow key to advance directly to the SCOPE 
MODE or press the RIGHT arrow key to step through the BLEACH, SCOPE and CURING MODES and press 
the SELECT button to enter that mode. Warning: Scope Mode is no longer used. Do not stay in scope mode.
Follow all other instructions for operation of that mode.
Press SELECT to stay in the CURE mode and the LEFT or RIGHT arrow keys to alternate between the two 
curing timers C1 and C2.

Using the new Sapphire Plus Curing Handpieces
The curing handpiece for the new Sapphire Plus light has been designed to work with existing Sapphire lights 
(reverse compatible). 
Follow all operating instructions for the selected handpiece. 

Install Power Cord, Install Handpiece, and Power Up
Install the AC power cord into the power inlet on the back of the Sapphire Plus light source and into a grounded 
AC socket.
Insert the proximal end of the fiber optic light guide on the curing / bleaching handpieces into the light guide 
receptacle on the front of the Sapphire Plus light source.
Turn the power switch on the back of the system to the ON ( I ) position. Press the SELECT button to enter the 
Curing mode (CURE) or use the LEFT or RIGHT arrow keys to choose the BLEACH modes and press SELECT 
to enter that mode.
Warning: Double check that the system is in the right mode as indicated by the green indicator light.
The Sapphire Plus is ready for use. It is necessary to read and understand all usage and safety precautions 
before use of this system.

9

Содержание Sapphire plus

Страница 1: ...Dental Curing Light with Optional Tooth Whitening Conforms to AAMI Std ES60601 1 IEC Stds60601 1 6 80601 2 60 62366 Certified to CSA Std C22 2 NO 60601 1 6 Instructions for Use...

Страница 2: ...nel 7 Accessories 7 Set up and Operation Instructions 8 Light Meter Radiometer 9 Curing Mode 9 10 Curing Instructions 10 Bleach Whitening Mode 11 12 Whitening Instructions 12 15 Sanitizing and Sterili...

Страница 3: ...placement must be made by a factory authorized representative Lamp replacement and fuse replacement must be done by a qualified service technician or by factory authorized representative Do not touch...

Страница 4: ...e or tip Do not autoclave the light shield provided with the unit Do not autoclave the light guide Removable curing tip is autoclavable Do not use abrasives or liquid chemicals to clean the tip or the...

Страница 5: ...erfere with other medical devices Since there is no guarantee in a particular installation that other device that interference will not occur test the installation prior to use Ensure that other devic...

Страница 6: ...OR SINGLE TOOTH WHITENING IF THE LIGHT OUTPUT IS GREATER THAN 3000 mW cm2 Call DenMat and return the instrument for repair 4 Do not saturate the controller keypad area with disinfectant or other liqui...

Страница 7: ...sor recognizes that they have been inserted and that the xenon arc lamp can be used when needed Control panels for curing bleaching are present The user interfaces offer a wide range of timing increme...

Страница 8: ...ct button B Timer display C Activator button Each handpiece has an activator button to activate the lamp 6 Removable Autoclavable Tip Optical device for exact placement of light 7 Lamp Access Plate Ac...

Страница 9: ...information Light Guide and Tips The Light Guide provided with the Sapphire Plus provides an optical path for transmitting light energy from the source to the tips Both the Light Guide and removable t...

Страница 10: ...ow key to advance directly to the SCOPE MODE or press the RIGHT arrow key to step through the BLEACH SCOPE and CURING MODES and press the SELECT button to enter that mode Warning Scope Mode is no long...

Страница 11: ...be set from 1 to 10 seconds It is recommended that multiple 5 second exposures be used for resins that need a little more time WARNING Single exposures longer than 5 seconds must be limited to use wit...

Страница 12: ...light cable and any curing tips which will be used Ensure they are in good working order Check to make sure there are no burrs or sharp edges or features which could cut a patient or the user of the...

Страница 13: ...dicated Warnings and Precautions start the system After the timing cycle is completed and the light turns off test the bottom surface for hardness using a dental probe If the material is not completel...

Страница 14: ...n of this user s guide Start the lamp by pressing the ACTIVATOR button on the handpiece or the START STOP button on the control panel Stop the lamp by pressing the ACTIVATOR button or START STOP again...

Страница 15: ...dish Use the brush applicator to apply a 1 1 5 mm layer onto the labial surface of the teeth 9 Install a sanitized Sapphire Whitening Crystal in place of the curing tip on the end of the Sapphire Plu...

Страница 16: ...or any cracking or lifting before re applying the whitening gel If the seal has broken from the tissue it is recommended to remove and replace the Paint On Dental Dam and begin with Step 4 17 At the e...

Страница 17: ...mile White In Office 25 with Take Home 22 Combo WHT0030 LumiBrite Chairside Whitening System 035359155 LumiBrite Desensitizing Gel 034159555 Paint On Dental Dam Kit Blue 404210 Paint On Dental Dam Syr...

Страница 18: ...lus System WARNING Unplug the power cord prior to cleaning with any liquid Allow the system to cool prior to cleaning Avoid spilling any liquid on the system Do not use flammable liquids to clean the...

Страница 19: ...he way Rotate the light guide 1 4 to 1 2 turn in the light guide receptacle Look for the display to light up on the handpiece If the display on the handpiece responds try starting the light again with...

Страница 20: ...de 5 Unplug the lamp from the 2 pin lamp socket To Insert New Lamp 1 Orient the lamp facing forward in the lamp guides on the chassis 2 Push toward the top of the unit until the lamp slides securely o...

Страница 21: ...pair at 1 800 433 6628 Transportation When transporting the Sapphire Plus remove the power cord and the light guide from the Sapphire Plus light source Hold the handle or carry it with both hands one...

Страница 22: ...will be evaluated for the cause and extent of failure or damage by Den Mat Holdings LLC Service Representatives Subject to the results of this evaluation Den Mat Holdings LLC will authorize warranty r...

Страница 23: ...posite veneers Core Paste XP Syringe 030654110 The recognized standard in core build up materials in a fast auto mixing syringe that can be easily syringed directly into canals Infinity Syringeable Va...

Страница 24: ...dings LLC All rights reserved 823057820 R 3 07 18CM EC REP MDSS GmbH Schiffgraben 41 30175 Hannover Germany Symbols Glossary available at www denmat com symbols Den Mat Holdings LLC 1017 W Central Ave...

Отзывы: