31 MAX MASS LIMIT 0
This limit applies to all possible mass entries. Exceeding this limit prevents a mass from entry. The
default limit is 1.5g
30 MIN MASS LIMIT 0
This limit applies to all possible mass entries. An entry below this limit will be rejected, the default is
0.2g.
21 MISFIRE LIMIT 1
If the sample is not igniting then it is called a misfire. The unit checks the temperature rise 30 seconds
after firing, and if the rise is below the limit a misfire is declared and the process is terminated. The
default value is 0.44C. The misfire limit checking can be disabled by entering zero.
18
TIME AND DATE
The CAL3K units have a date and time circuit which keeps the time. The data and time are attached to
every calibration, result and event. The following command set the time and date:
4 TIMESET HH:MM:SS 1
This command sets the time in HH:MM:SS. The internal clock runs for about 1 week without power and
then the time (and date) must be set.
5 DATESET YY:MM:DD 1
This sets the date of the CAL3K internal real time clock. Note that the clock remains life up to a week
without power. The entry is the form:
YY/MM/DD
The date can be used as a Group Identification (GID). This is not available on the
CAL3K-S.
122 DATE IS GID 1
This YES/NO entry allows for replacing the GID with the DATE in the form YY/MM/DD. The GID (Group
Identification) is attached to every sample recording. Note that the first 8 characters of the GID are left
blank.
The time can be used as a Sample Identification (SID). This is not available on the
CAL3K-S
. When this
feature is enabled then the auto-increment of the SID is disabled.
121 TIME IS SID 1