DDS CALORIMETERS OPERATIONS MANUAL V1.1
2022
Website:
www.ddscalorimeters.com
44
Digital Data Systems (Pty) Ltd.
CAL3K Advanced Bomb Calorimeters
www.ddscalorimeters.com
The actual data output starts 30 seconds before firing in the initial period and ends after the main
period. A single line of data are printed every 6 seconds as the data are gathered (in Real Time).
Four data output formats are available:
WXYZ.
The RT output is enabled by:
61 SET RT FORM WXYZ 1
This command enables and sets the REAL TIME output format to WXYZ. Selecting anything but WXYZ
disables the real time output. The WXYZ format can be tested by printing the header with CMD=53.
53 PRT RTIME HEADER 0
The header (Data descriptor) is printed. If the SET RTHEAD EN CP, CMD=60 has a selection (C=Customer,
P=Parameter) then this info is printed as well.
60 SET RTHEAD EN CP 1
The REAL TIME output (from 30 seconds before firing) can be preceded by Customer (C) and Parameters
(P). If none is desired enter anything but CP.
If the RT printout is enabled it may be advisable to disable the normal result print.
86 SET RESULT ENABL 1
This command enables (YES) or disables (NO) the spontaneous printing of results after a determination.
The results are entered in memory and can be printed later by any of the print commands.
Enter ‘NO’ to disable the result print.
25
CLEARING
It is possible that the CAL3K memory becomes useless because too many items are clogging up the
effective search and storage. It is therefore advisable to clear the memory once every 1 to 6 month
depending on usage.
25.1
CLEAR RESULT MEMORY
39 CLEAR RESULTS 1
This command clears all results. Depending on the unit the CAL3K keeps from 300 to 700 results.
Enter ‘YES’ to clear the result memory.