vPad-XSC for Impulse 4000 USER MANUAL
// "4. Press 'Next Step' button after defib has charged." | medium
//xtest "DISCHARGE defib!\n\nVerify energy is 43 to 58J" | DEHL,VFB1,50 | 5
//
//// Repeat for 100J energy (high range test)
//prompt "DEFIB ENERGY TEST\n\n"\+
// "1. Set defib energy to 100 JOULES.\n\n"\+
// "2. Charge up the defib.\n\n"\+
// "3. Wait two seconds.\n\n"\+
// "4. Press 'Next Step' button after defib has charged." | medium
//xtest "DISCHARGE defib!\n\nVerify energy is 85 to 115J" | DEHL,VFB1,100 | 2
//
//// Repeat for 200J energy (high range test)
//prompt "DEFIB ENERGY TEST\n\n"\+
// "1. Set defib energy to 200 JOULES.\n\n"\+
// "2. Charge up the defib.\n\n"\+
// "3. Wait two seconds.\n\n"\+
// "4. Press 'Next Step' button after defib has charged." | medium
//xtest "DISCHARGE defib!\n\nVerify energy is 170 to 230J" | DEHL,VFB1,200 | 2
//
//// Change automatic charge time limit to 10 seconds
//// Pass/fail is by manual entry because energy setting is not specified
xctrl "Setting Charge Time limit to 10 sec maximum..." | lim2,max,10 | 5
//
// Charge time test at max energy (high range test)
prompt "CHARGE TIME TEST\n\n"\+
"1. Set defib energy to operate from battery.\n\n"\+
"2. Set defib energy to MAXIMUM.\n\n"\+
"3. DO NOT CHARGE until Impulse 4000 shows 'MAXESTART'.\n\n"\+
"4. Discharge IMMEDIATELY AFTER defib reaches set energy.\n\n"\+
"5. Press 'Next Step' button when ready to proceed..." | medium
xtest "Charge when Impulse 4000 shows MAXESTART, then discharge.\n\n"\+
"Verify charge time is less than 10 seconds" | DCHS,VFB1,300
// Set Phase 3 ECG simulator to Coarse AFIB
//xctrl "Setting ECG to Coarse AFIB..." | ECGDA10 | -2
//
// Enable automatic assessment for sync delay (20 - 64 milliseconds)
xctrl "Setting sync delay range to 20 - 64 msec..." | lim3,range,20,64 | 2
// Cardioversion test (high range test; sync to Q-wave)
prompt "CARDIOVERSION TEST\n\n"\+
"1. Set defib energy to 20 JOULES.\n\n"\+
"2. Enable defib SYNC mode.\n\n"\+
"3. Charge up the defib.\n\n"\+
"4. Wait two seconds.\n\n"\+
"5. Press 'Next Step' button after defib has charged." | medium
xtest "DISCHARGE defib!\n\nVerify delay is 20 to 64 msec" | DSL,AF2,20 | 5
xctrl "Go to LOCAL mode" | #LOC
prompt "End of Impulse 4000 demo"
//power off
// End of script
Appendix A/Chapter 4
#