vPad-XSC for Impulse 4000 USER MANUAL
3.2.3 ECG Simulation
Impulse 4000
simulates ECG signals while conducting defibrillator and pacemaker tests. A Checklist may
select an ECG waveform through the Interface. This is typically done prior to performing a defib or pacer
test. ECG configuration statements are of the form
xctrl “<
info
>" | ECG,<
lead
>,<
wave
>
where:
< info >
is a description of the waveform, which is briefly displayed to the user
< lead >
is the ECG lead configuration,
< wave >
is the ECG waveform identifier.
3.2.3.1 Example 1: Configuring ECG Simulation
The following vPL statement configures
Impulse 4000
's ECG simulator:
xctrl “Setting ECG to NSB120" | ECG,II,NSB120
For this statement,
ECG
is the
Impulse 4000
remote command for ECG simulation,
II
lead configuration, either I or II
NSB120
is the waveform designation
This type of statement would typically precede a statement that tests defibrillator energy or other
aspect of defibrillator performance. Note that only certain waveforms should be used with
Defibrillator tests
(*)
or Cardioversion tests
(**).
ECG waveform designations:
ZERO
zero output, ECG off
NSB30
Normal sinus rhythm @30BPM
SQU
2-Hz square wave
NSB60 **
Normal sinus rhythm @60BPM
PUL
4-second pulse
NSB80 **
Normal sinus rhythm @80BPM
SIN0.O5
0.05-Hz sine wave
NSB120 **
Normal sinus rhythm @120BPM
SIN0.5
0.5-Hz sine wave
NSB160
Normal sinus rhythm @160BPM
SIN1
1-Hz sine wave
NSB200
Normal sinus rhythm @200BPM
SIN10
10-Hz sine wave
NSB240
Normal sinus rhythm @240BPM
Operation/Chapter 3
#
Page 11