// Prompt user to prepare pacer for the test
prompt "PACER PULSE TEST\n\n"\+
"1. Set pacer unit to READY.\n\n"\+
"2. Set rate to 70 PPM, output to 30 mA.\n\n"\+
"3. Set mode to FIXED.\n\n"\+
"4. Start pacing, then press 'Next Step'." | medium
// Get IP4k to perform pacer pulse test
xtest
"WAIT!\n\nVerify current is 27 to 33 mA\n"\+
"at 68 to 70 PPM" | PPPN,30,70,20
// Repeat for 120 mA, 140 PPM output
// First 2 results are discarded
//prompt "PACER PULSE TEST\n\n"\+
// "1. Set rate to 140 PPM, output to 120 mA.\n\n"\+
// "2. Start pacing, then press 'Next Step'." | medium
// xtest
"WAIT!\n\nVerify current is 108 to 132 mA\n"\+
// "at 138 to 142 PPM" | PPP2,120,140 | 2
//
// Enable automatic assessment for refractory test
xctrl "Setting paced refractory limit to 200 - 500 ms..." | lim8,range,200,500 | 2
xctrl "Setting sensed refractory limit to 500 ms maximum..." | lim9,max,500 | 2
//
// Prompt user to prepare pacer for refractory test
prompt "PACER REFRACTORY TEST\n\n"\+
"1. Connect pacer ECG leads to Impulse 4000.\n\n"\+
"2. Set rate to 140 PPM, output to 30 mA.\n\n"\+
"3. Set mode to DEMAND.\n\n"\+
"4. Start pacing, then press 'Next Step'." | medium
xtest
"WAIT!\n\nVerify results are within displayed limits" | PRFS
xctrl "Go to LOCAL mode" | #LOC
prompt "End of Impulse 4000 pacer demo"
// End of script