vPad-XSC for Impulse 4000 USER MANUAL
3.2.2 Resetting Limits to Defaults
The following vPL statement resets all limit settings to defaults:
xctrl “<
info
>" | rslm
where:
< info >
is a descriptive message which is briefly displayed to the user.
Table 2
lists the limit default values, which are assigned at power-on of the Interface, or in response to the
"rslm"
command as above.
TABLE 2: Default Limit Settings
TEST LIMIT
ID
DEFAULT VALUES
LIMIT TYPE
PARAMETER
1
PARAMETER
2
defib energy
1
target
±
15
%
percent
defib charge time
2
max
15
s
-
defib sync delay
3
range
20
ms min
65
ms max
pacer pulse current
4
target
±
10
%
percent
pacer pulse rate
5
target
±
2
PPM
units
pacer pulse width
6
none
-
-
N/A
7
none
-
-
pacer paced refractory period
8
none
-
-
pacer sensed refractory period
9
none
-
-
Operation/Chapter 3
#
Page 10