vPad-XSC for Impulse 4000 USER MANUAL
3.3 Defibrillator Tests
Once a Checklist has initialized communication with the Interface, configured
Impulse 4000
test settings
and, optionally, prepared limits for automatic pass/fail assessment, vPad is ready to execute a range of
defibrillator tests. A typical Checklist will include prompts instructing the operator how to set up the
defibrillator and
Impulse 4000
, followed by remote test statements.
While executing a test, the Interface acts as an intermediary that relays commands to
Impulse 4000
and
reformats the numerical results into a form suitable for vPad-Check and the Test Record.
vPL statements which trigger a remote test begin with the
“xtest”
keyword, in the form
xtest “<
instr
>" | <
cmd
>,<
par1
>,<
par2
>,...
where:
< instr >
is an instruction that is displayed to the user while the test is in progress,
< cmd >
is a command representing the test to be executed by
Impulse 4000
, and
< par1 >
< par2 >
... are optional parameters associated with automatic pass/fail assessment of the
measurement results, separated by commas (“
,
”).
The number of parameters specified in the
“xtest”
statement depends on the test to be executed.
Automatic pass/fail assessment is disabled if a required parameter has not been specified in the statement.
In this case, a "---" will appear in the Test Record report.
For a detailed description of the vPL
“xtest”
statement, refer to
vPad-Check Operating Manual
.
Operation/Chapter 3
#
Page 13